Service Procedure 13c: Replacing the ECS Wire Harness
After removing the cassette’s plastic bottom panel and its main housing (as per the technical
manual):
Remove the contact box (see Fig. 49), and remove the clear plastic cable guard from the
cassette’s trolley using an 8 mm (5/16”) socket.
Remove the wire harness cable from under the loosened cable guard.
Remove the topshell housing to get access to the unit’s frame.
Unplug the ECS circuit board connector. You can now dispose of the cable.
Connect the new ECS wire harness cable to the ECS circuit board.
After replacing the upper plastic housing, thread the cable back into the cable guard on
the trolley, and reinstall the screw.
Reinstall the ECS contact box, and the top shell. Make sure that the unit’s “antistatic” wire is
oriented so that it will make contact with the track.
Fig. 50
22