37
Guidance and Manufacturer’s declaration - Electromagnetic Immunity
Guidance and manufacturer’s declaration - electromagnetic immunity
Immunity Test
IEC 60601 Test Level Compliance Level
Electromagnetic Environment Guid-
ance
Conducted RF
IEC 61000-4-6
Radiated RF
IEC 61000-4-3
3 Vrms 150 kHz to 80
MHz
3 Vrms 80 kHz to 2,5
GHz
3V
3 Vm
Portable and mobile RF
communications equipment should
be used no closer to any part of the
Maxi Air
system
including cables,
than the recommended separation
distance calculated from the
equation applicable to the frequency
of the transmitter.
Recommended separation distance
d=1,2
√
P
d=1,2
√
P 80 MHz to 800 MHz
d=2,3
√
P 800 MHz to 2.5 GHz
Where P is the maximum output
power rating of the transmitter in
watts (W) according to the
transmitter manufacturer and d is the
recommended separation distance in
meters (m)
Field strengths from fixed RF
transmitters, as determined by an
electromagnetic site survey, a)
should be less than the compliance
level in each frequency range. b)
Interference may occur in the
vicinity of equipment marked with
following symbol:
NOTE 1
At 80 MHz and 800 MHz, the higher frequency range applies.
NOTE 2
These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorp-
tion and reflection from structures, objects and people.
Содержание Maxi Air
Страница 1: ... with people in mind 04 MS 01_7GB July 2013 Maxi Air INSTRUCTIONS FOR USE ...
Страница 9: ...9 Intentionally left blank ...
Страница 11: ...11 14 17 11 10 16 18 9 15 12 13 1 2 3 4 5 6 7 8 ...
Страница 15: ...15 Intentionally left blank ...
Страница 29: ...29 Intentionally left blank ...
Страница 31: ...31 Data label LATEX REF Maxi Air MAS014000 WW LOT Patient Name Date L ...