35
Electromagnetic Compatibility
Medical electrical equipment needs special precautions regarding EMC (Electromagnetic compatibility) and needs to
be used according to the EMC information below. The
Maxi Air
system is intended for use in the electromagnetic
environment specified below. The customer or the user of the
Maxi Air
system should make sure that it is used in such
an environment.
WARNING
Stacking or placing other electrical equipment next to this device is not recommended, it can
interfere with the equipment’s operation and safety. Portable and mobile radio-frequency (RF)
communications equipment can interfere with this equipment operation and safety.
WARNING
The equipment may cause radio interference or may disrupt the operation of nearby equip-
ment. It may be necessary to take action, such as reorienting, relocating the equipment or
shielding the location.
Guidance and manufacturer’s declaration - electromagnetic emissions
Emissions test
Compliance
Electromagnetic environment - guidance
RF emissions
CISPR 11
Group 1
The
Maxi Air
system
uses RF energy only for its internal function.
Therefor, its RF emissions are very low and are not likely to cause
any interference in nearby electronic equipment. RF emissions.
RF emissions
CISPR 11
Harmonic emissions
IEC 61000-3-2
Voltage fluctuations/
flicker emissions
IEC-61000-3-3
Class B
Class B
Complies
The
Maxi Air
system
is suitable for use in all establishments other
than domestic and those directly connected to the public low
voltage power supply network that supplies buildings used for
domestic purposes.
Содержание Maxi Air
Страница 1: ... with people in mind 04 MS 01_7GB July 2013 Maxi Air INSTRUCTIONS FOR USE ...
Страница 9: ...9 Intentionally left blank ...
Страница 11: ...11 14 17 11 10 16 18 9 15 12 13 1 2 3 4 5 6 7 8 ...
Страница 15: ...15 Intentionally left blank ...
Страница 29: ...29 Intentionally left blank ...
Страница 31: ...31 Data label LATEX REF Maxi Air MAS014000 WW LOT Patient Name Date L ...