5-5
Auto Program Selection Guidelines
MANUAL AIR ADJUSTMENT
Use Manual Air Adjustment to individually select air pressures for each set of cushions when Auto Program Air Adjustment
does not provide appropriate patient positioning.
1.
Press AIR ADJUST button. The LED flashes when changes can be made.
2.
Press the INFLATE / DEFLATE arrows for the corresponding region, adjust target pressures as needed.
NOTE:
Allow sufficient time for cushions to respond to air pressure adjustments.
3.
Check patient for proper alignment.
•
Allow a cradling effect, approximate 40% depression into cushions where patient lies.
•
Perform a hand check beneath head, shoulders, hips and heels:
•
Slide hand under Cover Sheet and beneath cushions to rest on Cushion Support Surface.
•
Lift hand slightly to verify approximately 4 inches of air is supporting the patient.
4.
Press AIR ADJUST button to exit Air Adjust.
4’10 in
4’11 in
5’0 in
5’1 in
5’2 in
5’3 in
5’4 in
5’5 in
5’6 in
5’7 in
5’8 in
5’9 in
5’10 in
5’11 in
6’0 in
6’1 in
6’2 in
6’3 in
6’4 in
6’5 in
6’6 in
6’7 in
6’8 in
<130 lb
<133 lb
<136 lb
<139 lb
<141 lb
<143 lb
<145 lb
<148 lb
<151 lb
<154 lb
<157 lb
<160 lb
<163 lb
<166 lb
<170 lb
<174 lb
<178 lb
<182 lb
<187 lb
<193 lb
<197 lb
<203 lb
<208 lb
>130 lb
>133 lb
>136 lb
>139 lb
>141 lb
>143 lb
>145 lb
>148 lb
>151 lb
>154 lb
>157 lb
>160 lb
>163 lb
>166 lb
>170 lb
>174 lb
>178 lb
>182 lb
>187 lb
>193 lb
>197 lb
>203 lb
>208 lb
SHORT
HEIGHT
MEN
TALL
<108 lb
<110 lb
<112 lb
<114 lb
<116 lb
<118 lb
<120 lb
<123 lb
<125 lb
<127 lb
<131 lb
<133 lb
<136 lb
<139 lb
<142 lb
<146 lb
<149 lb
<152 lb
<155 lb
<159 lb
<163 lb
<167 lb
<173 lb
SHORT
HEIGHT
WOMEN
TALL
Содержание KinAir MedSurg Pulse
Страница 1: ...KinAir MedSurg Pulse USER MANUAL P N 409100 AH Rev B 06 2018 with people in mind ...
Страница 2: ......
Страница 40: ...6 3 Grounded Equipment Power OFF Power ON O I ...
Страница 43: ......