Battery Pack in Use
The battery pack will automatically take over powering the pump whenever the AC
cord is removed from the wall outlet.
When running under battery power, a battery symbol appears in the top right hand
corner of the display.
In order to maximize battery life, the power alarm indicators will not
illuminate when operating the AC600 under battery power. You may verify
therapy is, or is not, being provided via the LCD screen.
The amount of charge remaining in the battery pack is indicated by the black block
inside the battery symbol.
When the battery has only 5% of the charge remaining, the battery symbol will flash
and the remaining charge will be displayed as a percentage. Recharge the battery as
soon as possible.
Then, when the battery has 1% charge, the audible alarm will sound but the pump
will continue to run. When the battery has no charge left the pump will shut down
completely. Recharge the battery.
After use, disconnect the battery from the pump to maintain battery charge level. If
the pump is put into standby mode it will beep after 10 minutes of inactivity,
reminding you to either disconnect the battery or to plug the pump into an electrical
outlet.