15
Checking the Status of the Battery
Battery Label
To Check the Battery Status:
1. Remove the battery from the pump unit.
2. Press the Reset/Status Latch which will activate the 2 LEDs on the top of the
battery.
If the battery pack is not on charge, the LEDs will display one of the following
conditions:
LED 1
LED 2
Battery Status
Green
Green
The Battery Pack is OK. It has > 80% charge.
Green
Green
The Battery Pack is OK. It has between 30-80% charge.
Green
The Battery Pack is OK. It has < 30% charge.
Red
Green
The Battery Pack needs recycling*. It has > 30% charge.
Red
The Battery Pack needs recycling*. It has < 30% charge.
Red
The Battery Pack has a fault.
The Battery Pack has a fault.
LED 1
LED 2
80-100% Charge
Battery Pack OK
Battery Pack
needs recycling
Battery Pack
Fault
30-80% Charge
<30% Charge
Press Reset / Status Latch
Jack Socket Polarity