background image

(iii)

General Safety

Before you connect the system pump to a mains socket, read carefully all the 
installation instructions in Section 4 - System Set Up. The system has been designed 
to comply with regulatory safety standards including:

• UL60601-1 & CAN/CSA C22.2 No 601-1

Precautions

For your own safety and the safety of the equipment, always take the following 
precautions:

• Regularly check the system when in use, for pump operation and garment fit.

• Keep the pump away from sources of liquids and do not immerse in water.

• Do not expose the system, especially the garments, to naked flames, such as 

cigarettes etc.

• Do not store the system in direct sunlight.

• Switch off the electrical supply to the pump by disconnecting the pump from the 

mains socket before cleaning and inspection.

• Do not use hypercarbonate or phenolic based cleaning solutions.

• Ensure the system is clean and dry prior to storage.

• Only the pump and garment combination as indicated by Huntleigh Healthcare 

should be used. The correct function of the product cannot be guaranteed if the 
incorrect pump/garment combinations are used.

Safety Warnings

• Electrical equipment may be hazardous if misused. The pump's 

case back should only be removed by authorised technical 
personnel.

• Do not use the pump in the presence of flammable gases such 

as anaesthetic agents.

• Make sure that the mains power cable is positioned to avoid 

causing a trip or other hazard, and is clear of moving bed 
mechanisms or other possible entrapment areas.

Caution

Electromagnetic compatibility (EMC).  This product complies with
the requirements of applicable EMC Standards.  The use of
accessories not specified by the manufacturer may result in
increased emissions by, or decreased immunity of, the equipment,
affecting its performance

Содержание 507003

Страница 1: ...Intermittent Pneumatic Compression System User Manual Flowtron Universal ...

Страница 2: ......

Страница 3: ...ions 4 System Operation 5 To Use One Garment 7 To Use Different Garments on Each Limb 8 Settings Adjustment 9 Troubleshooting 10 Options 11 Battery Operation 12 Connecting the Battery Pack 12 Battery Pack in Use 13 Checking the Status of the Battery 14 Recharging the Battery Pack 15 Alternative Method of Charging 16 Warranty and Service 17 Cleaning Instructions 19 Ordering Information 20 Technical...

Страница 4: ... ii ...

Страница 5: ...ot use hypercarbonate or phenolic based cleaning solutions Ensure the system is clean and dry prior to storage Only the pump and garment combination as indicated by Huntleigh Healthcare should be used The correct function of the product cannot be guaranteed if the incorrect pump garment combinations are used Safety Warnings Electrical equipment may be hazardous if misused The pump s case back shou...

Страница 6: ... iv ...

Страница 7: ...rly clot formation The system comprises of a pump that can be used in conjunction with an extensive range of Huntleigh Healthcare inflatable garments including foot calf and thigh length versions The system is designed to be simple and easy to operate The pump automatically adjusts to the correct therapy profile depending upon which garment type foot calf or thigh length is connected An optional b...

Страница 8: ... Depending on the garment type used other clinical applications are also appropriate The FG foot garment in particular has a wide range of clinical applications Full details for clinical applications are included in the packaging of every garment The type of garment used on an individual patient must be specified by a physician ...

Страница 9: ...ysician notified 4 When used for DVT prevention continuous external pneumatic compression is recommended until the patient is fully ambulatory Uninterrupted use of the Flowtron Universal system is encouraged 5 The Flowtron Universal system should be USED WITH CAUTION on patients with Insensitive extremities Diabetes Impaired circulation Fragile or impaired skin Contraindications The Flowtron Unive...

Страница 10: ...imbs should be checked during every shift and more often if the patient has known circulatory or skin problems or is diabetic Note Many patients are at risk for pressure ulcers on the heel Use of the foot garments does not negate the necessity for heel protection and proper skin care Clinical judgment should be used to determine if the patient s skin condition requires additional measures such as ...

Страница 11: ...uired for use Apply the prescribed Huntleigh Healthcare garment to the patient by carefully following the instructions included in the garment packaging Connect the garments to the pump air tubing and ensure the connectors click into place Run Standby Button LCD Display and Function Buttons Air Feed Tube Connectors Power Alarm Indicators ...

Страница 12: ...ll begin this example shows the pump running and the left garment inflated to 40 mmHg Refer to Battery Operation on page 13 when operating the pump under battery power TO STOP press the green RUN button a second time The green lights will go out and therapy will stop Garment Type Inflation Pressure mmHg Inflation Hold seconds Cycle seconds Foot garments e g FG100 200 130 3 30 Single chamber garmen...

Страница 13: ...check that the garments remain correctly fitted to the patient To Use One Garment When the pump is in standby page 5 system operation connect one garment to either tubing outlet The display will confirm that only one garment is connected If a garment is disconnected while the pump is running the system will alarm Lo If the single garment option is still required simply press the RUN button to rese...

Страница 14: ...e to use different garment types on each limb if prescribed by the patient s physician Simply connect the garments to either tubing outlet and the correct therapy cycle will be administered Check on the display panel to confirm the garment types connected ...

Страница 15: ...ommended therapy for each garment type and does not require any setting by the clinician or nurse If the physician requires different therapy settings limited changes can be made through a service link Contact your local Huntleigh sales representative for details ...

Страница 16: ...w the pump to run until it senses a normal inflation it will then reset itself Recommended Checks To confirm that the fault has corrected watch the display for approximately one minute If the fault has not been corrected a fault message will flash during that time In addition the pump will refuse to start if it does not recognize the attached garment Pump Display Problem Corrective Action Lo Garme...

Страница 17: ...t either high or low volume setting Press the button to confirm The selected volume setting will remain in place during subsequent uses of the pump Patient Hours Meter Press the button Use the arrow buttons to select Hours Run Press the button to display the total hours run Press the button again the pump will revert to its previous state To Clear the Patient Hours Meter Press the button Use the a...

Страница 18: ...cent to Options Use the arrow buttons to highlight Language Press the button to confirm Press the button to confirm Use the arrow buttons to scroll through the available languages Press the button to confirm Use the arrow buttons to highlight Exit Press the button to confirm ...

Страница 19: ... 8 hours of operation can be expected from a fully charged battery pack Connecting the Battery Pack From the front of the pump engage the pack on the runners and slide into place until a click is heard To remove the pack depress the latch at the rear of the unit and slide pack forward and out Reset Status Latch Battery Pack ...

Страница 20: ...ack is indicated by the black block inside the battery symbol When the battery has only 5 of the charge remaining the battery symbol will flash and the remaining charge will be displayed as a percentage Recharge the battery as soon as possible Then when the battery has 1 charge the audible alarm will sound but the pump will continue to run When the battery has no charge left the pump will shut dow...

Страница 21: ... Battery Status Green Green The Battery Pack is OK It has 80 charge Green Green The Battery Pack is OK It has between 30 80 charge Green The Battery Pack is OK It has 30 charge Red Green The Battery Pack needs recycling It has 30 charge Red The Battery Pack needs recycling It has 30 charge Red The Battery Pack has a fault The Battery Pack has a fault LED 1 LED 2 80 100 Charge Battery Pack OK Batte...

Страница 22: ...lt condition Recharging the Battery Pack The battery pack is automatically recharged whenever the pump is connected to an AC outlet On the display the arrow confirms the pump is connected to AC power and the battery is charging LED 1 LED 2 Battery Status Green Green The Battery Pack is OK It is fully charged Green The Battery Pack is charging Red Green The Battery Pack needs recycling The Battery ...

Страница 23: ...ugging it into the battery charger as shown below Avoid charging the battery near source of heat or in direct sunlight Caution Only Huntleigh approved battery chargers must be used Use of a non approved charger may result in damage to the battery pack or ineffective charging and therefore voids the warranty ...

Страница 24: ...or any other reasons shall be at Huntleigh Healthcare s sole option limited to i the refund of the full purchase price for such Products plus the invoice shipping charges or ii replacement thereof free of charge to the Customer with Products conforming to the then current specifications thereof The parties agree that these remedies shall be exclusive HUNTLEIGH HEALTHCARE AND ITS AFFILIATES SHALL H...

Страница 25: ...on Universal pump should receive annual preventative maintenance All required servicing should be performed by Huntleigh Healthcare service personnel Please obtain a Returned Goods Authorization from the Customer Service Department prior to returning the product to Huntleigh Healthcare Display the RGA number prominently on the side of the package used to return the product This will insure prompt ...

Страница 26: ...ol determined by the customer to be valid However using certain chemicals and cleaners will cause severe damage to the pump Never use phenolic hypocarbonate based disinfectants containing cationic surfactants ammonia based compounds or antiseptic solutions such as Steriscol or Hibiscrub Do not spray cleaning solutions directly onto the pump The pump cannot be sterilized All garments are single pat...

Страница 27: ...ontact Customer Service Department Huntleigh Healthcare L L C 40 Christopher Way Eatontown NJ 07724 3327 Telephone 800 223 1218 Product Code Description Package AC600 Flowtron Universal Pump 1 each BBP600 Battery Pack 1 each ABC600 Battery Charger 1 each ...

Страница 28: ... of in a dustbin Recycle Recycle USA Electrical Safety Standards Tested to UL60601 1 Classified by Underwriters Laboritories Inc with respect to electric shock fire and mechanical hazards only in accordance with UL60601 1 25EA CAN CSA C22 2 No 601 1 M90 Classified by Underwriters Laboritories Inc with respect to electric shock fire and mechanical and other specified hazards only in accordance with...

Страница 29: ...Double Insulated Type BF Run Standby Refer to the user manual Fuse SN Serial number Ref Model number Do not dispose of in a dustbin Recycle refuse Refer to accompanying documents Dangerous voltage With respect to electric shock fire and mechanical hazards only in accordance with UL60601 1 and CAN CSA C22 2 No 601 1 i ...

Страница 30: ...Pa Storage Pump Temperature range 40 F to 158 F 40 C to 70 C Relative humidity 10 to 100 non condensing Atmospheric pressure 500 hPa to 1060 hPa Storage Battery Pack Temperature range 14 F to 86 F 10 C to 30 C Relative humidity 30 to 75 Atmospheric pressure 500 hPa to 1060 hPa Environmental Protection Please dispose of this unit in accordance with local regulations ...

Страница 31: ... Pack Information Recharge time for fully discharged battery 5 hours on standby Recharge time with optional wall charger 5 hours End of battery life indicators 500 recharge cycles resulting in flashing RED LED indicators ...

Страница 32: ...26 ...

Страница 33: ...27 ...

Страница 34: ...28 ...

Страница 35: ...e Ltd Unit 6 38 Eaglehurst Road Ellerslie Auckland New Zealand T 64 9 525 2488 F 64 9 525 2433 Deutschland HNE Healthcare GmbH Im Hülsenfeld 19 D 40721 Hilden T 49 0 2103 97 11 0 F 49 0 2103 97 11 80 South Africa Huntleigh Africa Pty Ltd Willem Cruywagen Avenue Stand 120 Klerksoord Pretoria South Africa T 27 12 542 4680 F 27 12 542 4982 Danmark Huntleigh Healthcare A S Vassingerødvej 52 DK 3540 Ly...

Страница 36: ... Huntleigh Healthcare Limited 2002 507900US_05 ...

Отзывы: