Page | 28
Commercial in confidence. User Manual Revision CTK-REV7-14082019
9. TROUBLE SHOOTING
Symptom
Possible Cause
Action
All actuators fail to work
Power disconnected. or
batteries discharged
Ensure mains supply lead
is connected
Control box and handset
plugs disconnected
Check the plugs are fully
inserted in the control box
and any lock out devices
fitted to the bed. Check that
plugs are fully inserted in
the sockets on the deck
frame. Check that the
handset controls are not
deactivated on the patient
lockout facility.
Blown fuse
Contact your local Arjo
Service Department.
Damaged handset cable
Replace handset
One actuator fails to work
Actuator plug disconnected Check that actuator plug is
fully inserted in control box.
Check that the handset
controls are not deactivated
on the patient lockout
facility if fitted.
Main lift actuators fail to
work, and an audible
beeping sound can be
heard when handset is
activated
Control box requires
calibrating possibly due to
bed being operated when
batteries were fully
discharged
Ensure mains power lead is
connected and carry out
calibration procedure.
Brakes or steering lock not
effective
Wear or damage
Adjust or replace castor
Backrest CPR release not
effective
Release mechanism not
correctly adjusted
Check and re-adjust
Содержание Prioma
Страница 1: ...04 PA 00 EN_03 08 2019 INSTRUCTIONS FOR USE Prioma Medical Bed ...
Страница 32: ...Page 31 Commercial in confidence User Manual Revision CTK REV7 14082019 12 PRIOMA BED LABELS AND WARNINGS ...
Страница 35: ...Page 34 Commercial in confidence User Manual Revision CTK REV7 14082019 Intentionally left blank ...
Страница 36: ...Page 35 Commercial in confidence User Manual Revision CTK REV7 14082019 Intentionally left blank ...
Страница 37: ...Page 36 Commercial in confidence User Manual Revision CTK REV7 14082019 Intentionally left blank ...