8. MAINTENANCE
The following preventative maintenance procedures should be carried out every 12
months.
NOTE:
Other than general maintenance activities all other service or repair work must
be performed by qualified and trained persons who are approved by Arjo.
WARNING
The bed must be disconnected from the mains power supply before starting any
maintenance activity.
GENERAL
1
IF FITTED with a bed lock out device set all functions to locked.
2
Examine the bed for obvious signs of damage. All aspects of the equipment
should operate as intended. Check that all nuts, bolts and other fasteners are tight
and are not missing.
3
Examine flexible cables and conduits for cuts, cracks, abrasions or other
deterioration.
4
Check that the power supply plug is not damaged. If either the power cable or
plug is damaged, then both the cable and plug must be replaced as a complete
assembly by an approved service agent.
BRAKE LOCKING SYSTEM
1
Apply the brakes and push the bed forwards. If any of the four braking castors
rotate the brake is not fully effective.
2
(Central Locking) Check the two steering lock castors for correct operation. If
either the brakes or the steering locks are not working correctly contact an
approved service agent.
Содержание Prioma
Страница 1: ...04 PA 00 EN_03 08 2019 INSTRUCTIONS FOR USE Prioma Medical Bed ...
Страница 32: ...Page 31 Commercial in confidence User Manual Revision CTK REV7 14082019 12 PRIOMA BED LABELS AND WARNINGS ...
Страница 35: ...Page 34 Commercial in confidence User Manual Revision CTK REV7 14082019 Intentionally left blank ...
Страница 36: ...Page 35 Commercial in confidence User Manual Revision CTK REV7 14082019 Intentionally left blank ...
Страница 37: ...Page 36 Commercial in confidence User Manual Revision CTK REV7 14082019 Intentionally left blank ...