Page | 9
Commercial in confidence. User Manual Revision CTK-REV7-14082019
4. ELECTRONIC FUNCTIONS DESCRIBED
- Continued
NOTE:
This control function is not available on the Patient Handset and Inset
Patient Control Panel. It is available on all other control devices fitted to the
bed.
To move the mattress platform up or down, press and hold the appropriate button
or buttons on the control device. The beds platform height can be adjusted
through a range
(depending on castor choice)
of 370mm low height to a high
height of 820mm.
The mattress frame height adjustment uses two actuators working in conjunction
and separate to each other creating the vertical lift and enabling the bed to tilt in
opposite directions.
(3) Knee Positioning Adjustment
NOTE:
This control function is available on all control devices fitted to the bed.
To move the knee positioning section up and down, press and hold the
appropriate button or buttons on the control device. The thigh section is
adjustable through a range of 0° to 35°, which provides adjustable support to the
patient’s legs.
(4) Auto Contour Positioning
NOTE:
This control function is not available on the Attendant Control Panel
(ACP). It is available on all other control devices fitted to the bed.
To move the auto contour positioning press and hold the appropriate button or
buttons on the control device. Auto Contour Positioning provides synchronised
movement of the backrest and knee positioning in one operation. The position
helps provide a more natural resting environment for the patient.
(5) Tilt Positioning – Trendelenburg / Reverse Trendelenburg
CAUTION
Trendelenburg function should only be used under professional supervision.
NOTE:
This control function is not available on the Patient Handset and Inset
Patient Control Panel. It is available on all other control devices fitted to the
bed.
To move the tilt positioning press and hole the appropriate button or buttons on
the control device. The tilt button allows the bed to be positioned in a foot or head
down tilt position. Tilt positioning achievable is between 0° and 14° for both head
and foot down tilting.
Содержание Prioma
Страница 1: ...04 PA 00 EN_03 08 2019 INSTRUCTIONS FOR USE Prioma Medical Bed ...
Страница 32: ...Page 31 Commercial in confidence User Manual Revision CTK REV7 14082019 12 PRIOMA BED LABELS AND WARNINGS ...
Страница 35: ...Page 34 Commercial in confidence User Manual Revision CTK REV7 14082019 Intentionally left blank ...
Страница 36: ...Page 35 Commercial in confidence User Manual Revision CTK REV7 14082019 Intentionally left blank ...
Страница 37: ...Page 36 Commercial in confidence User Manual Revision CTK REV7 14082019 Intentionally left blank ...