![Arjo Prioma Скачать руководство пользователя страница 22](http://html1.mh-extra.com/html/arjo/prioma/prioma_instructions-for-use-manual_2972288022.webp)
Page | 21
Commercial in confidence. User Manual Revision CTK-REV7-14082019
4. GENERAL BED OPERATION
- Continued
3/4 Length Fold Down Tubular Rail Safety Lock Indicator
The side rail locating pin has a yellow coating that allows for visual inspection of the
locking device. When the side rail is in the raised position, viewing the release knob from
the side there should be no yellow shaft visible. If there is yellow shaft visible the safety
side is
NOT
securely locked in the upright position and needs to be adjusted.
The yellow coating should
NOT
be visible from above when the safety side is correctly
locked in position.
Side rail release and lock knob shown in unlocked and locked positions
WARNING
•
After adjustment of the side rail visually check the locking pin location. There
should be no yellow shaft on the locking pin visible if locked correctly.
•
DO NOT use other manufacturer’s side rails on the Prioma bed.
Содержание Prioma