Page | 15
Commercial in confidence. Information for Use Manual Revision COU-SE-REV12-220820
7. PRIOMA - SEMI ELECTRIC COUCH SPECIFICATIONS
Dimensions
Couch Model
HC1
HC2
HC3
Number of Sections
1 section
2 section
3 section
Length (mm) (+/-10mm)
1945
1945
1945
Width (mm) (+/-10mm)
700
700
700
Width with Fold-down Arms (+/-10mm)
755
755
755
Deck Height Range (mm) (+/-10mm)
460 to 935
460 to 935
460 to 935
Backrest Angle Adjustment (+/-5°)
N/A
75°
75°
Knee Positioning Angle (+/-5°)
N/A
N/A
-30°
Safe Working Load & Product Weight
(approximate)
Couch (without accessories)
76 Kg
84 Kg
89 Kg
With Fold-down Arms Fitted
+12.0 Kg
+8.0 Kg
+8.0 Kg
With Fold-down Arms & Paper Roll
Holder fitted (excluding paper roll)
+13.5 Kg
+9.5 Kg
+9.5 Kg
Safe Working Load SWL
200 Kg
200 Kg
200 Kg
Electrical Data
Power In
100 - 240V AC 50/60Hz, Max current 2.5A
Duty Rating
Intermittent 10%, Max 2 min / 18 min
Electric Shock Protection
Class 2 Type B
Liquid Ingress Protection
IPX6 Waterproof
Standards
This product has been designed and manufactured to comply with International
Standard IEC 60601-2-52 were applicable.
WARNING
The clinically trained person should consider the patients weight, size and mobility to
establish the safe use of the Prioma couch.