Commercial in confidence. Information for Use Manual Revision COU-SE-REV12-220820 

1.  WARNINGS & CAUTIONS

 - 

Continued

 

 

GENERAL SAFETY WARNINGS 

  This product is not designed to be used with a mattress. 

  Before operating the couch make sure that the patient is safely positioned to 

avoid entrapment. 

  Before transporting the couch make sure the power cable is disconnected from 

mains power and stored safely. Adjust the couch to a suitable height to move 
without stooping, or straining. 

  Always retract 

“castor use mode” when the couch is stationary. 

  When the patient's condition could lead to patient entrapment the upholstered 

platform should be left in the flat position whilst unattended. 

  The clinically qualified person responsible should consider the age, size and 

condition of the patient before using the couch and its accessories. 

  Always ensure that all hospital lines and power cables are free from possible 

entanglement with mechanical parts of the couch. 

  Always ensure that attendant and patient limbs are free from potential 

entrapment areas before operating any functions. 

  It is recommended that the product installation process is carried out by two 

people. 

  It is recommended to place the couch on a clean, flat floor or platform. 

 

WARNING 

  Ensure that the patient, caregivers and other equipment and furniture are not 

positioned to become either trapped or crushed during couch adjustment. 

  Ensure that the power lead is not being stretched. 

  Ensure that the power cable cannot become entangled with moving parts of 

the couch. 

 

DISCONNECT 

the electrical power cable from the mains and store before 

transporting the couch. 

  If the power cable shows any sign of damage, replace immediately. 

  Some functions may cause injury to the patient if unintentional movement 

occurs. 

CAUTION

 

Do not operate the controls for more than the control systems prescribed duty 

cycle (no more than 2 minutes within an 18 minute period of time).

 

 

 
 

 

Содержание Prioma HC1

Страница 1: ...INSTRUCTIONS FOR USE COU SE REV12 220820 Semi Electric Prioma Couch Range...

Страница 2: ...nee Positioning Adjustment 8 5 4 Castor Fixed Positioning or Mobile Mode 9 5 5 Fixed Leg Adjustment 9 5 6 Fold down Arms if fitted 10 5 7 Paper Roll Holder if fitted 11 6 PRODUCT CARE 11 6 1 Cleaning...

Страница 3: ...in the presence of flammable gases Do not use electrically operated couches in operating theatres Use the couch to support one patient only The Prioma couch is intended for one person use and may beco...

Страница 4: ...ital lines and power cables are free from possible entanglement with mechanical parts of the couch Always ensure that attendant and patient limbs are free from potential entrapment areas before operat...

Страница 5: ...ponents are made from PP and ABS plastic 2 2 Features Standard Electrically adjustable height 75mm diameter retractable wheels for uncompromised stability Levelling foot for uneven floors Manually ope...

Страница 6: ...Product Overview A Backrest Breath Hole if fitted B Upholstered Backrest C Upholstered Seat D Upholstered Footrest E Adjustable Leg F Wheels x 4 G Fixed Leg x 3 H Retracting wheels activation pedal I...

Страница 7: ...not stretched kinked or crushed Do not allow the power supply cord to trail on the floor where it may cause a trip hazard Make sure the power supply cord does not become entangled with moving parts of...

Страница 8: ...lding fixtures and equipment when operating couch functions The height adjustment of the couch is driven by an electric motor that is activated by a foot operated controller Height adjustment occurs o...

Страница 9: ...spring locking mechanism allowing the backrest to be raised or lowered Release the handle when the desired position is reached to lock in position 5 3 Knee Positioning Adjustment 3 Section Model Only...

Страница 10: ...l levers are angled below the horizontal position the four castors are lowered raising the supporting legs off the ground The couch is now mobile CAUTION Disconnect the bed power cable from the source...

Страница 11: ...Prioma 1 Section Arm Length 1000mm O A Fig 4 To Lower Grip and hold the Arm in the middle of the top rail A and pull up as indicated by the directional arrow fig 1 to unlock Controlling the Arm allow...

Страница 12: ...processes on the Prioma Couch The following procedure also applies to the accessories 6 1 Cleaning Clean all surfaces with a disposable cloth soaked in a neutral detergent and hand hot water Start by...

Страница 13: ...n both the cable and plug must be replaced as a complete assembly by an approved service agent CASTORS FIXED MOBILE MODE 1 Step and push down on the lever pedals Push the couch forward checking that e...

Страница 14: ...he damage to maintenance staff 2 Visual inspection of the Cable and Cable Insulation 2 1 Inspect the cable for any nicks or cuts in the insulation 2 2 Inspect for any evidence that the cable may have...

Страница 15: ...ootswitch plugs disconnected Check the plugs are fully inserted in the control box Blown fuse Contact your local Arjo Service Department Damaged footswitch cable Replace footswitch Control box overhea...

Страница 16: ...oximate Couch without accessories 76 Kg 84 Kg 89 Kg With Fold down Arms Fitted 12 0 Kg 8 0 Kg 8 0 Kg With Fold down Arms Paper Roll Holder fitted excluding paper roll 13 5 Kg 9 5 Kg 9 5 Kg Safe Workin...

Страница 17: ...cation Label Safe Working Load Refer to Instructions for Use Date of Manufacture Product Reference Number 11 Digit Code Representing Various Options Build Types Product Serial Number Manufacturer Indi...

Страница 18: ...any attempt be made to open the device WARNING DANGER OF EXPLOSION DO NOT HEAT OR INCINERATE High pressure gas The sudden release of gas at high pressure could cause serious injury or death Use suitab...

Страница 19: ...pressure 70 kPa To 106 kPa 11 WARRANTY AND SERVICE Arjo s standard terms and conditions apply to all sales A copy is available on request These contain full details of warranty terms and do not limit...

Страница 20: ...ensure ergonomic patient handling personal hygiene disinfection diagnostics and the effective prevention of pressure ulcers and venous thromboembolism we help professionals across care environments t...

Отзывы: