background image

18 

TROUBLESHOOTING

Basic troubleshooting procedures for the MaxxAir ETS Mattress Replacement System are 

provided in the following charts. Each chart deals with a specific symptom and provides 

a symptom / possible cause / remedy approach to use in identifying the solution. The 

symptom will describe the condition that the unit is in. The possible cause will describe 

several likely reasons for the symptom and the steps to take in verifying the cause. The 

Remedy will describe the final solution to the symptom and cause.

SYMPTOMS, POSSIBLE CAUSE AND REMEDIES

 

SYMPTOM

POSSIBLE CAUSE

REMEDY

Cells do not inflate.

Hose may not be properly 

connected to mattress or Air 

Unit.

Turn Expansion knob to 

proper position.
Properly connect hoses.

Expansion Side Bolster 

does not expand.

Bolster may not be properly 

connected to hose set.

Properly connect hose 

to bolster.

Cells do not inflate.

Air Supply Unit power 

switch is on but LED does 

not illuminate.

Air Supply Unit may be 

controlled by wall switch.
Power cord may be unplugged 

from wall outlet.
Circuit breaker or fuse for wall 

outlet may be tripped or blown.
Air Supply Unit may be 

damaged.

Plug Air Supply Unit into 

different wall outlet.
Verify Power cord is 

plugged in.
Reset circuit breaker or 

replace fuse.
Contact Arjo Service 

Representative.

One or more cells do not 

inflate.

Cell may not be properly 

connected to hose set. Cells 

may be damaged.
Cell fabric may be twisted 

above connector, preventing air 

from entering cushion.

Properly connect hose 

to cell.
Replace cell.
Adjust cell and verify 

hose connection.

Air Supply Unit is turning 

on and off, or varying 

speed.

Air Supply Unit may be 

controlled by a wall switch.

Too many devices may be in 

operation or plugged into same 

circuit as Air Supply Unit.

Air filter may be clogged with 

dust and debris.

Plug Air Supply Unit into 

different wall outlet.

Unplug or turn off one 

or more devices.

Clean or replace air 

filter.

Air Supply Unit air flow 

increases or decreases 

without adjustment and 

does not respond to Soft / 

Firm button.

PC Board in Air Supply Unit 

may be damaged.

Contact Arjo Service 

Representative.

Содержание MaxxAir ETS

Страница 1: ...MaxxAir ETS INSTRUCTIONS FOR USE 310115 AH Rev 3 09 2019 Mattress Replacement System...

Страница 2: ...us improvement we reserve the right to modify designs without prior notice The content of this publication may not be copied either whole or in part without the consent of Arjo WARNING To avoid injury...

Страница 3: ...OPERATION 15 CARE AND CLEANING 16 GENERAL INSTRUCTIONS 16 DAILY CARE AND CLEANING 16 WEEKLY CARE AND CLEANING 17 CLEANING PROCEDURE 17 TROUBLESHOOTING 18 SYMPTOMS POSSIBLE CAUSE AND REMEDIES 18 SPECI...

Страница 4: ...4 Intentionally left blank...

Страница 5: ...forth in the written limited warranty included with this product Information in this publication may be subject to change at any time Contact Arjo for updates IMPORTANT INFORMATION FOR USERS In order...

Страница 6: ...6 Intentionally left blank...

Страница 7: ...peed blower that provides adjustable air pressure A variable cycle time turning valve is used to activate the turn assist function The Turn Assist feature slowly inflates the appropriate turning bladd...

Страница 8: ...sition Verify wheels are locked before any patient transfer to or from the bed Bed Height To minimize risk of falls or injury the bed should always be in the lowest practical position when the patient...

Страница 9: ...ty rules and institution protocols concerning patient and caregiver safety Power Cord Position power cord to avoid a tripping hazard and or damage to the cord Ensure power cord is kept free from all p...

Страница 10: ...ssories and see the Questions and Information section of this guide for contact information Transfer Board If a mechanical lift device is not available a transfer board or device should always be used...

Страница 11: ...e 36 in wide bed frame the expansion knob should be in line with 36 in When using the 42 in wide bed frame the expansion knob should be in line with 42 in decal as shown Patient left side surface will...

Страница 12: ...Push INSTAFLATE button to deactivate the fast inflation Raise and lock side rails and make sure any positioning aids are properly in place Adjust bed height to lowest level AIR PRESSURE ADJUSTMENTS 1...

Страница 13: ...button to enable mattress to turn right and back to center and adjust turn time from 3 30 minutes Press LEFT TURN Button to activate left hold Cycle time will display zero zero 00 Press the CYCLE butt...

Страница 14: ...where incontinence and drainage occur Ensure linens under patient are not wrinkled INCONTINENCE DRAINAGE Use moisture impermeable underpads for incontinent patients Wipe surface clean and replace bed...

Страница 15: ...15 GENERAL OPERATION Avoid contact of sharp instruments with the MaxxAir ETS MRS Punctures cuts and tears may prevent proper inflation and air pressure maintenance...

Страница 16: ...at the intervals described All nylon fabric with moisture vapor permeable backing cover sheets must be laundered and disinfected after each patient use prior to use by a new patient Contact Arjo for i...

Страница 17: ...l level cleaner disinfectant mixed to manufacturer s specifications to the surface until thoroughly wet Scrub all areas of therapy unit hoses bed frame mattress and if used the mattress inserts side r...

Страница 18: ...y Unit may be controlled by wall switch Power cord may be unplugged from wall outlet Circuit breaker or fuse for wall outlet may be tripped or blown Air Supply Unit may be damaged Plug Air Supply Unit...

Страница 19: ...or replace air filter Contact Arjo Service Representative Blower still runs but cannot change between modes i e cannot change LED lights from INSTAFLATE to STATIC CPU needs to be reset If you turn mas...

Страница 20: ...Polyurethane 0 35 mm Cover Material Nylon fabric with moisture vapor permeable backing Base Material Nylon Laminated PVC Air Supply Unit Size Width 16 5 in W 42 cm Depth 6 7 in D 17 cm Height 10 5 in...

Страница 21: ...0601 1 2 Second Edition EN 5501 1998 Group 1 Class B EN 61000 3 2 Class A EN 61000 3 3 CLASSIFICATION Classification in accordance with UL 60601 1 CSA C22 2 No 601 1 and IEC 60601 1 Class II Type BF I...

Страница 22: ...E mail Italy promo arjo com MIDDLE EAST Arjo Middle East FZ LLC Of ce 908 9th Floor HQ Building North Tower Dubai Science Park Al Barsha South P O Box 11488 Dubai United Arab Emirates Direct 971 487 4...

Страница 23: ...23 Intentionally left blank...

Страница 24: ...ergonomic patient handling personal hygiene disinfection diagnostics and the effective prevention of pressure ulcers and venous thromboembolism we help professionals across care environments to conti...

Отзывы: