Troubleshooting
001.1
1720.EN rev
. 13
Automatic KWIKtrak Turntable
Manual KWIKtrak Turntable
PROBLEMS
TO CHECK
The turntable does not move
when the lift is brought on its
track.
• Check if the lift is on.
• Make sure that there is contact between the contact blades of the lift
and the contact plates of the contact box.
• Check if batteries are not completely discharged.
• Check if contact plates are clean and free of oxidation. If needed,
wash them with a normal detergent.
The turntable does not pivot
even if electrical feeding is
OK.
• Make sure the Quick-release knob is in “Powered Rotation” position.
• If the turntable does not work as intended after checking the previous
point, immediately contact your Arjo agent.
The turntable stops moving
between track and beeps.
The turntable is in over current condition.
• Verify if an object is restraining the turntable from rotating.
• If no object is blocking the rotation and the turntable does not recover
by itself, use the Quick-release to complete patient transfer and call
your Arjo agent.
PROBLEMS
TO CHECK
The turntable does not move
when the cord is pulled.
• Check if the turntable moves when the other cord is pulled.
• Check that the lift is centered onto the pivoting track, and call your
Arjo agent.
The central pivoting track
does not lock in position.
• Verify the position of the activation arm, which must be in the “up”
position.
• If the activation arm does not move up in position, call your Arjo
agent.
Troubleshooting