11
Care and Maintenance
001.1
1720.EN rev
.13
Preventive Maintenance Schedule
The product is subjected to wear and tear, and the following maintenance instructions must be acted upon
when specified to ensure that the unit remains within its original manufacturing specifications. Care and
maintenance must be carried out in accordance with the preventive maintenance schedule below.
User Inspections
Annual Inspection by an Authorized Technician
The KWIKtrak Turntable must be inspected by a qualified Arjo technician on an annual basis to ensure the
proper function of its components, as well as the safety of the patients and caregivers that use the
product. Please contact your local agent to make arrangements.
Cleaning
To clean the KWIKtrak Turntable, wipe it down with a damp cloth using warm water and mild detergent.
To ensure a better rolling surface for the trolley wheels, clean the inside of the track every 4 months. To do
so, insert a damp cloth in the opening and slide it from one end of the track to the other.
WARNING: Safety related maintenance and authorized service must be carried out by qualified
staff, fully trained in servicing procedures by Arjo, and equipped with correct tools and proper
documents including Parts List and Service Manual. Failure to meet these requirements could
result in personal injuries and/or unsafe equipment.
WARNING: The maintenance described in the following checklist is the minimum that the
manufacturer recommends. In some cases more frequent inspections should be carried out.
Continuing to use this product without conducting regular inspections or when a fault is found
will seriously compromise the safety of the user/patient. Service and preventive maintenance can
be arranged with the manufacturer. Preventive maintenance specified in this manual can prevent
accidents and reduce repair costs.
WARNING: Never proceed to maintenance or service while product is in use with a patient.
Inspections for the KWIKtrak Turntable
Frequency
Initially
Before
every use
Every
month
Inspect for evidence of external damage or missing parts.
X
X
Make sure that the activation stations are clean.
X
Make sure the lift moves smoothly between the turntable track and fixed
tracks (tracks are well aligned).
X
Make sure end stoppers of the turntable are in place.
X
Clean the contact plates of the contact box with a damp cloth.
X
Inspect pull cords for wear.
X
CAUTION:
Do not use phenol, chlorine or any other type of solvent that may damage the finish.