16
Electromagnetic Compliance
001.1
1720.EN rev
. 13
Electromagnetic Immunity
Guidance and Manufacturer’s Declaration -
Electromagnetic Immunity - For all Equipment and Systems
The KWIKtrak Turntable is intended for use in electromagnetic environment specified below. The
customer or the user of the KWIKtrak Turntable should assure that it is used in such an environment.
Immunity test
IEC 60601 test level
Compliance
level
Electromagnetic environment -
guidance
Electrostatic
discharge (ESD)
IEC 61000-4-2
±8 kV contact
±15 kV air
±8 kV contact
±15 kV air
Floors should be wood, concrete or
ceramic tile. If floors are covered with
synthetic material, the relative
humidity should be at least 30%.
Electrical fast
transient/burst
IEC 61000-4-4
Not applicable
Not applicable
Mains power quality should be that of
a typical commercial or hospital
environment.
Surge
IEC 61000-4-5
Not applicable
Not applicable
Mains power quality should be that of
a typical commercial or hospital
environment.
Voltage dips,
short
interruptions and
voltage
variations on
power supply
input lines
IEC 61000-4-11
Not applicable
Not applicable
Mains power quality should be that of
a typical commercial or hospital
environment. If the user of the
KWIKtrak Turntable requires
continued operation during power
mains interruptions, it is
recommended that the KWIKtrak
Turntable be powered from an
uninterruptible power supply or a
battery.
Power frequency
(50/60 Hz)
magnetic field
IEC 61000-4-8
30 A/m
50/60 Hz
30 A/m
50/60 Hz
Power frequency magnetic fields
should be at levels characteristic of a
typical location in a typical
commercials or hospital environment.