AZI Customer Service 800-528-7411 or 602-470-1414
Page 36 of 57
7.
CALIBRATION
Factory Calibration Service
The Jerome
®
J605's gold film sensor is inherently stable and does not require frequent calibration.
The interval between calibrations depends upon the application and frequency of use; however,
the recommended interval is every 12 months.
The Jerome
®
J605 has been factory calibrated using laboratory equipment containing NIST
traceable permeation tubes. In order to calibrate the Jerome
®
J605, a sophisticated calibration
system is required that ensures stability of the calibration gas source, eliminates any pressure in
the calibration gas stream, and controls the temperature of the calibration environment.
We strongly recommend you take advantage of our calibration and maintenance service at
Arizona Instrument. Service includes filter replacement, component testing, and instrument
calibration to NIST traceable standards. A certificate of calibration is issued from AZI when your
instrument is factory calibrated.
For scheduling and shipping authorization, call AZI Customer Service at
(800) 528-7411 or (602) 470-1414, or visit our website at
Verification of Functionality and Quality Control
The Functional Test Module (AZI P/N: Z2600 0918 or Z2600 0930) is used to determine if your
instrument is functioning correctly between recommended annual factory calibrations. It allows
you to have complete confidence in the sample results. This test verifies proper instrument
operation through the introduction of a known concentration of hydrogen sulfide into the Jerome
®
analyzer.
THIS TEST DOES NOT CALIBRATE THE INSTRUMENT.
If your application requires frequent verification of instrument function, this test demonstrates the
instrument’s operation and function. Recording FTM results in an instrument log provides a
quality control/quality assurance record of instrument function between regular calibrations. If
test results fall within the expected range, you may assume the instrument is functioning
correctly.
See
APPENDIX B - J605 FUNCTIONAL TEST MODULE
on page 55 for more information
about the FTM procedures. Complete instructions for use are supplied with the test kit, AZI P/N
Z2600 0918.
To order the FTM, contact your AZI Sales Representative at (800) 528-7411 or (602) 470-1414.