AZI Customer Service 800-528-7411 or 602-470-1414
Page 12 of 57
5.
INSTRUMENT OPERATION
J605 Main Screen Display
The J605 Main Screen is depicted below, followed by explanations of the labeled items.
SYSTEM TIME
– As set in the SYSTEM menu. (See
BATTERY
– Indicates current charge level and charging status (See
SOUND ICON
– Indicates if the audible alarm is turned on or muted, as set in the SYSTEM
SYSTEM DATE
– As set in the SYSTEM menu. (See
SENSOR SATURATION METER
– Graphically indicates the hydrogen sulfide saturation level
of the J605 sensor. (See
USB ICON
– Indicates proper functioning of USB ports, and flashes during USB transmission of
data. (See
). If the USB icon is not present, this option is not
activated. Contact your AZI Sales Representative for details on the USB option.
RANGE
– Indicates in which Range the J605 is set to sample: 0, 1, 2 or A (Auto-range), as set in
the SAMPLE menu. (See Range in
AUTO SAMPLING INDICATOR
– Indicates if the J605 is set to sample automatically at
regular intervals, as set in Sample Mode in the SAMPLE menu. (See
HYDROGEN SULFIDE READING
– The hydrogen sulfide concentration detected by the
J605, in the units indicated.
UNITS
– Units for the hydrogen sulfide reading: parts per million (ppm) or parts per billion
(ppb), depending on the Range in use and the sample concentration detected. Readings of
zero are displayed as 0.000 ppm or 0.00 ppb, depending on the Range being used.
AVAILABLE KEYPAD COMMANDS
– The bottom line of the screen indicates the currently
available keypad commands. From the main screen, as shown here, the Main Menu can
be accessed using the ► key.