AZI Customer Service 800-528-7411 or 602-470-1414
Page 21 of 57
VIEW SENSOR STATS
Use
VIEW SENSOR STATS
to display the current saturation level of the sensor, the number of
regenerations performed on the sensor, and the total number of samples read by the sensor. Press
any key to exit the screen when finished.
SET SYSTEM TIME
The J605 maintains the system time using a 24-
hr clock format. Use the ▲and ▼arrow keys to
change the value in the field where the cursor is flashing. Use the ◄ and ► arrow keys to move
between the two digits within the hour or minute fields, and use the
A
and
B
keys to switch
between the hour and minute fields. Use
ENTER/START
to save the indicated time as the
system time or
ESC
to exit without saving.
SET SYSTEM DATE
Use the ▲and ▼arrow keys to change the highlighted field, and the ◄ and ► arrow keys to
move between characters within the day or year fields. Use the
A
and
B
keys to move between
the day, month, and year fields. Press
ENTER/START
to save the system date or
ESC
to exit
without saving.
RUN DIAGNOSTICS
Selecting
RUN DIAGNOSTICS
will start the J605
’
s self-diagnostic procedure. As the
instrument proceeds through each diagnostic check,
follow the on-screen prompts
, and the
instrument will display the specific check being performed and the pass/fail result of the check. If
the check is not applicable to the current configuration of the instrument, the result field will be
blank instead of indicating pass or fail.
•
After viewing each result, press any key (except I/O or ESC) to move to the next diagnostic
check. (I/O should not be used during the diagnostics because I/O will power off the
instrument and ESC should only be used when specifically indicated by the instrument.)
•
When testing the keypad, follow the on-screen prompts to press each of the keys in turn. The
J605 will not indicate a pass/fail for this test. Instead, if all of the keys work and the
instrument proceeds to the next diagnostic test, the keypad diagnostic has passed.
•
If the 4-20mA output function of the J605 is not in use, press ENTER/START to cycle
through the 4, 8, 12, 16 and 20mA checks in the diagnostic routine. If the 4-20mA output is
in use, the diagnostics can be used to verify proper operation. Refer to
After the last diagnostic result has been displayed, the instrument will display “Done,” and
pressing any key will return to the main screen of the J605.
SET CONTRAST
The display contrast can be set from 0 (highest contrast) to 127 (least contrast) by u
sing ▲ and ▼
to adjust the contrast. By default, ▲ and ▼will adjust the contrast in increments of 1 unit. Press
B
to switch to a coarse adjustment of 10 units for each press of ▲ or ▼. Press
A
to switch back
to the default fine adjustment mode. Press
ENTER/START
to save the contrast setting and exit.
CHANGE PASSWORD
Select
CHANGE PASSWORD
to change the password on the SYSTEM menu. Use
▲ and ▼
to change each
alphanumeric character, and ► and ◄ to move between characters. Press
ENTER/START
to proceed after entering the password.