Timed Regeneration
If the unit is to be operated unattended for extended periods, AZI recommends that the sensor be
regenerated regularly. Operation under JCS or Data Logger control automatically regenerates
saturated sensors. The Option Functionality can control regeneration on a regular basis, every 6,
12 or 24 hours.
The regeneration intervals are set through a combination of switch settings as shown in the
following table:
------- SW100---------
DIP 1 DIP 2
REGENERATION
Interval (Hrs.)
OFF
OFF
OFF
ON
OFF
6
OFF
ON
12
ON
ON
24
Auto-Sample
If a Data Logger is either not connected (on Non-SMV units), not activated (on SMV units) or is
operating in the manual sampling mode, the following automatic sampling rates may be selected
with SW100 DIP switch settings:
SW100 DIP switch settings
Sampling frequency
DIP 3
DIP 4
DIP 5
ON
ON
ON
No automatic sampling
OFF
ON
ON
5 minutes
OFF
OFF
ON
15 minutes
OFF
ON
OFF
30 minutes
OFF
OFF
OFF
1 hour
The switches have no effect if a Data Logger is connected or activated and operating in
automatic sample mode as programmed through the JCS.