Syringe Technique
•
Pull and hold the syringe plunger
against the bar-stop.
•
Verify that the black mark on the
syringe plunger aligns with the 1cc
mark on the syringe barrel.
If it does not, the holder assembly
must be adjusted. Call AZI customer service at 602-281-
1745 or 800-528-7411 for assistance.
•
Insert the needle into the needle-guide of the bottle stopper.
•
Operate the plunger two or three times to pump mercury
vapor into the syringe. On the final stroke, pull and hold
the plunger against the bar-stop.
•
Holding the plunger against the bar-stop, remove the
syringe from the bottle and move it to the septum attached
to the instrument intake.
•
Continue to hold the plunger against the bar-stop and insert
the syringe needle into the septum.
•
Press “SAMPLE” on the instrument.