AZI Customer Service 800-528-7411 or 602-470-1414
Page 27 of 56
8.
JEROME
®
431-X TECHNICAL SPECIFICATIONS
Range
0.001 to 0.999 mg/m
3
Detection Limit
0.003 mg/m
3
Hg
Precision
5% relative standard deviation at 0.100 mg/m
3
Hg
Accuracy
±5% at 0.100 mg/m
3
Hg
Response time-sample mode
12 seconds
Response time-survey mode
3 seconds
Flow rate
750 ± 50ml/min (0.75 ± .05 liters/min)
Power requirements
100-120 VAC, 50/60 Hz, 1 A, or
220-240 VAC, 50/60 Hz, 1 A
Fuse
F1A 250V, 5mm X 20mm
Internal battery pack
- standard model
- XE model
Rechargeable Nickel Cadmium:
AZI P/N Z4000 0907
AZI P/N Z4000 0908
Operating environment
0º to 40 ºC, non-condensing, non-explosive
Case construction
Aluminum alloy
Dimensions – standard model
Dimensions – XE model
33 cm L x 15 cm W x 10 cm H (13" L x 6" W x 4" H)
35 cm L x 18 cm W x 18 cm H (14” L x 7” W x 7” H)
Weight – standard model
Weight – XE model
3.18 kilos (7 pounds)
3.5 kilos (8 pounds)
Digital meter display
Liquid crystal display (LCD)
Certification
CE mark on 220-240 V~, 431-XE model only.
Optional Communications Capability
Data output
1. RS-232 Serial, Baud Rate 1200 for use with Data
Logger, and/or Jerome
®
communication program.
2. RS-232 Serial data format with 0 & 20mA current
logic levels; Baud Rate 1200 (special industrial
applications) and Analog 20 mA output.
"With Option Functionality"
- See APPENDIX D - JEROME
Analog output
0 to 2V or 4 to 20 mA
Auto sample interval
5, 15, 30, or 60 minutes
Auto regeneration interval
6, 12 or 24 hours