5.
MAINTENANCE
Preventive Maintenance Calendar
To keep the Jerome
®
431-X operating at peak performance, follow the maintenance schedule
below as a guide. Since maintenance is more a function of application and amount of use rather
than time, your requirements may be different from the listed schedule. Call AZI Customer
Service at 800-528-7411, 602-470-1414, or e-mail [email protected] for additional guidance for
your environment and operation.
PART/COMPONENT
MAINTENANCE CYCLE
REFER TO
PAGE
Charge batteries
At least once per month, after 1 month's storage,
or when LO BAT appears
Page 17
Change .25 inch fritware
Weekly or as needed
Page 19
Change internal filters and
tubing
After 6 months of use or as needed
Page 20
Replace zero air filter
Annually
Factory calibration
Annually
Page 23
Functional check
Monthly or as needed
Appendix
A, Page 37
Replace batteries
Annually or as needed.
The battery pack contains NiCad batteries.
Dispose of properly.
Page 21
NOTE:
Install the zero air filter into the instrument's intake during storage.
1
C/M filters contain Sofnolime ™. For safety information see the supplier’s Safety Data Sheet (SDS) or call AZI
Customer Service at 800-528-7411, 602-470-1414, or e-mail [email protected] for a copy. Dispose of all filters
properly.
2
Zero air filters and scrubber filters contain Resisorb
®
or AddSorb
®
VA13. For safety information see the
supplier’s Safety Data Sheet (SDS) or call AZI Customer Service at 800-528-7411, 602-470-1414, or e-mail
[email protected] for a copy. Dispose of all filters properly.