8.3.
Calibration Menu
To get here, navigate:
[MENU]
CALIBRATION MENU
The calibration menu provides access to
the balance and heater calibration
routines.
8.3.1 Balance Calibration Menu
To get here, navigate:
[MENU]
CALIBRATION MENU
BALANCE CALIBRATION MENU
Span calibration is used when setting up
the instrument in a new location. The
balance linearity should be stable
throughout the balance’s life. Linearity
verification can be performed, if desired,
but only on the MAX
®
4000XL.
NOTE: From a cold start, allow the MAX
®
4000 to warm up for 60 minutes prior to
balance calibrations and verifications for optimum accuracy. The instrument
should not be moved, jarred, or otherwise disturbed in any way during the
calibration period. A vibration-free table will aid the calibration process. If
vibration is excessive, the instrument will stop the calibration process and
indicate the reason.