Aries HIGH PURITY WATER SYSTEM Скачать руководство пользователя страница 6

ARIES FilterWorks, a division of ResinTech, Inc. * 160 Cooper Road * West Berlin, NJ  08091

Phone:  (856) 768-9600 * Fax:  (856) 768-9601 * www.arieswater.com

3.0  GENERAL DESCRIPTION

This ultrapure water dispensing system has  been

engineered  to consistently produce Type I ultrapure

water.  The system is  user-friendly,  requires  minimal

space  and  maintenance,  and  harmonizes  with

existing  laboratory  equipment,  all  at  a  low  initial

investment and low operating cost.

A central dispensing station can supply the high

quality  water  needs  of  one  lab  bench  or  an  entire

laboratory.  The system may be free standing or wall

mounted where bench space is limited.

A  continuously  cycling  recirculation  pump

polishes the water assuring the highest quality water

is  always  available,  and  a  high  resolution  digital

monitor  is  built-in  for  continuous  display  of  water

purity.

A    timer  controller  operates  the  recirculation

pump,  solenoid  valve  and  resistivity  meter.    Bright

LED’s let you monitor overall system performance at

a glance.  The quick connect cartridge system makes

replacing filters a simple task. 

For  optimum  performance  the  Aries  High  Purity

Water  System  may  require  reverse  osmosis,

distillation,  or  deionizing  pretreatment.    The  feed

water  quality  will  determine  the  necessary

pretreatment, if any.   Optional pretreatment systems

are available to  operate from tap water that does  not

meet feed water specifications.

3.1 PRINCIPLES OF PURIFYING WATER

  Mixed  Resins  are  used  for  polishing  purity

levels from pre-treated, filtered, RO, distilled,

or DI water to a maximum of  18.17 megohms

resistivity.

  .2 micron final filter for Bacteria removal.

  Timer Controller runs system automatically for

30 minutes  every  2  hours  to  keep  bacterial

from blooming.

  WAKE  Button  interrupts  system  standby  to

start 30 minute run cycle.

    A  high  resolution  digital  purity  monitor

displays the ionic purity of the  treated  water

in Megohms to two decimal places.  The user

never has to guess  how  pure  the  water  is. 

The built-in pump will recirculate the water for

the highest available purity.

  Optional  Ultraviolet  Irradiation  is  used  for

organic reduction & bacterial reduction

  Optional .05 micron absolute filter for  pyrogen

reduction.

Содержание HIGH PURITY WATER SYSTEM

Страница 1: ...ARIES HIGH PURITY WATER SYSTEM Operation Maintenance Manual Version 1 1a...

Страница 2: ...u have any questions concerning the operation of the unit or need service refer to section 11 0 REQUESTING SERVICE When requesting service you will need the model and serial number as well as the purc...

Страница 3: ...Principles of Purifying Water 4 0 Product Specifications 5 0 Before Using Your Dispenser 5 1 Setup 5 2 Installing the Cartridges 5 3 Startup Procedure 5 4 Initial Polishing of new cartridges 6 0 Displ...

Страница 4: ...er cord grip the plug and pull it from the wall outlet NEVER PULL ON CORD DO NOT use this appliance outdoors This appliance is intended for commercial industrial institutional and professional use Use...

Страница 5: ...ure successful water system operation Trace organic chemicals can interfere with highly sensitive reactions and alter biological processes TDS Total Dissolved Solids conduct electricity This conductiv...

Страница 6: ...t a glance The quick connect cartridge system makes replacing filters a simple task For optimum performance the Aries High Purity Water System may require reverse osmosis distillation or deionizing pr...

Страница 7: ...sformer UV Combination for Bacteria and TOC Wall Mounting Kit Reverse Osmosis Pre treatment Packages Re circulating Spray Gun 0 05 micron Pyrogen removal filter Typical Applications Media Prep Tissue...

Страница 8: ...tion valve at this time 5 2 Installing the Cartridges There are three 3 cartridge positions The water flows from left to right Any pretreatment filters must be installed in position 1 furthest from th...

Страница 9: ...g to confirm the effectiveness of the treatment process is the owners responsibility 6 0 DISPLAY AND KEYPAD FUNCTIONS Be sure you understand the function of each button before using them Function Keys...

Страница 10: ...e top cover by removing the phillips head screws that attach it Gently lift the top cover up and left and set it aside Unclasp the clamps that hold the stainless steel body of the UV light Gently lift...

Страница 11: ...rom the outlet Place a wad of paper towels around each plunger to absorb the small amount water that will come out of the filters plungers and or upper fitting Remove the three cartridges starting wit...

Страница 12: ...s 1 5 0 micron prefilter 1 GAC carbon filter and 1 1 0 micron post filter 9 5 Aries HPL RO Replacement Membrane AM 12 7010 Replacement membrane is a thin film composite PA membrane with 100 GPD output...

Страница 13: ...se barb connection 9 12 Blank Sanitization Cartridges VP 17 0000 Blank VP Style cartridges for sanitization of unit 3 required per Aries System Sanitization kit sold separately 10 0 REQUESTING SERVICE...

Страница 14: ...turned on Bad Solenoid Have unit serviced by an authorized dealer or return to factory A filter cartridge is loaded with particles and plugged No flow Replace the filter cartridge Capsule filter is bo...

Страница 15: ...llege of American Pathologists ASTM American Society for Testing and Materials NCCLS National Committee for Clinical Laboratory Standards EU Endotoxin Units N L No Limit The measurements of pH in Type...

Страница 16: ...anty ARIES ResinTech Inc assumes no other responsibility or liability This warranty does not cover any product which has been subject to misuse neglect accident or used in violation of operating instr...

Отзывы: