7.2 Replacing the Filter Cartridges and Final Filter

The cartridges are expendable items which have a finite capacity to deionize and polish the water.   When the

digital display indicates the purity is less than your minimum requirements, the DI resin cartridges have

exhausted their capacity to perform.   Cartridge should be replaced 

every 6 months

 as a matter of routine

maintenance.   The final filter should be changed every time the cartridges are replaced

  Turn the unit off using the power switch.

  Open the outlet valve and leave open to purge any excess pressure.   Use a container to catch any

water that may splash from the outlet.

  Place a wad of paper towels around each plunger to absorb the small amount water that will come out

of the filters, plungers and/or upper fitting.

  Remove the three cartridges starting with Cartridge #3.   Push down on the cartridge gently and tilt

forward.   Quickly remove the filter and discard in the trash.   There will be some residual water inside the

filter.   Water will come out the outlet valve (still open) when the cartridge is depressed.

  Remove cartridge’s #2 and #1 in this order using the same procedure above.   When the plunger is

depressed some water will come out of the upper fitting next to the filter.

  Dry the unit and replace the cartridges as described in Section 5.2, Installing the Cartridge.

  The used cartridges may be disposed of in the same manner as any water filter cartridge.

  Replace the Final Filter with a new one each time the filter cartridges are replaced.

8.0  SANITIZATION

See Sanitization Kit for instruction.   Sold Separately.

Содержание HIGH PURITY WATER SYSTEM

Страница 1: ...ARIES HIGH PURITY WATER SYSTEM Operation Maintenance Manual Version 1 1a...

Страница 2: ...u have any questions concerning the operation of the unit or need service refer to section 11 0 REQUESTING SERVICE When requesting service you will need the model and serial number as well as the purc...

Страница 3: ...Principles of Purifying Water 4 0 Product Specifications 5 0 Before Using Your Dispenser 5 1 Setup 5 2 Installing the Cartridges 5 3 Startup Procedure 5 4 Initial Polishing of new cartridges 6 0 Displ...

Страница 4: ...er cord grip the plug and pull it from the wall outlet NEVER PULL ON CORD DO NOT use this appliance outdoors This appliance is intended for commercial industrial institutional and professional use Use...

Страница 5: ...ure successful water system operation Trace organic chemicals can interfere with highly sensitive reactions and alter biological processes TDS Total Dissolved Solids conduct electricity This conductiv...

Страница 6: ...t a glance The quick connect cartridge system makes replacing filters a simple task For optimum performance the Aries High Purity Water System may require reverse osmosis distillation or deionizing pr...

Страница 7: ...sformer UV Combination for Bacteria and TOC Wall Mounting Kit Reverse Osmosis Pre treatment Packages Re circulating Spray Gun 0 05 micron Pyrogen removal filter Typical Applications Media Prep Tissue...

Страница 8: ...tion valve at this time 5 2 Installing the Cartridges There are three 3 cartridge positions The water flows from left to right Any pretreatment filters must be installed in position 1 furthest from th...

Страница 9: ...g to confirm the effectiveness of the treatment process is the owners responsibility 6 0 DISPLAY AND KEYPAD FUNCTIONS Be sure you understand the function of each button before using them Function Keys...

Страница 10: ...e top cover by removing the phillips head screws that attach it Gently lift the top cover up and left and set it aside Unclasp the clamps that hold the stainless steel body of the UV light Gently lift...

Страница 11: ...rom the outlet Place a wad of paper towels around each plunger to absorb the small amount water that will come out of the filters plungers and or upper fitting Remove the three cartridges starting wit...

Страница 12: ...s 1 5 0 micron prefilter 1 GAC carbon filter and 1 1 0 micron post filter 9 5 Aries HPL RO Replacement Membrane AM 12 7010 Replacement membrane is a thin film composite PA membrane with 100 GPD output...

Страница 13: ...se barb connection 9 12 Blank Sanitization Cartridges VP 17 0000 Blank VP Style cartridges for sanitization of unit 3 required per Aries System Sanitization kit sold separately 10 0 REQUESTING SERVICE...

Страница 14: ...turned on Bad Solenoid Have unit serviced by an authorized dealer or return to factory A filter cartridge is loaded with particles and plugged No flow Replace the filter cartridge Capsule filter is bo...

Страница 15: ...llege of American Pathologists ASTM American Society for Testing and Materials NCCLS National Committee for Clinical Laboratory Standards EU Endotoxin Units N L No Limit The measurements of pH in Type...

Страница 16: ...anty ARIES ResinTech Inc assumes no other responsibility or liability This warranty does not cover any product which has been subject to misuse neglect accident or used in violation of operating instr...

Отзывы: