11.0  TROUBLESHOOTING GUIDE

SYMPTOM

CAUSE

REMEDY

Unit will not WAKE up

Wake  button  pushed  too

quickly.

WAKE button must be

pushed  at least _ second  to

initiate the polishing mode.

Display is off.

No power to the unit or unit in

Stand  By  mode.

Check the unit is plugged into

a live outlet.

Check the unit's fuse - replace

it if it is blown.

Press the Wake switch to

initiate the polish cycle

Unit blows fuses.

An electrical short exists in the

unit.

Return the unit to the factory

for repairs or have unit

serviced by an authorized

dealer.

Unit will not dispense water.

Supply water is not on.

Confirm that the supply water

is turned on.

Bad  Solenoid.

Have unit serviced by an

authorized dealer or return to

factory

A filter cartridge is loaded with

particles  and  plugged.

No flow

Replace the filter cartridge.

Capsule filter is bound with

air. Open bleed valve

Display indicates water purity

is below set point.

The supply water is improperly

treated.

Test the supply water TDS

add  additional  treatment.

The resin cartridges are not

seated.

Open  cartridge  housing  and

reseat  cartridge.

Resin is exhausted.

Install fresh cartridges.

Resin cartridges exhaust

quickly.

The supply water is improperly

treated.

Test the supply water TDS.

Add  additional  pretreatment

to  obtain  higher  quality.

Sub-micron filter plugs rapidly.

The supply water has colloids

and is improperly treated.

Perform a Silt Density Test.  If

it is higher than 1 add

additional  pretreatment.

                                                            

Содержание HIGH PURITY WATER SYSTEM

Страница 1: ...ARIES HIGH PURITY WATER SYSTEM Operation Maintenance Manual Version 1 1a...

Страница 2: ...u have any questions concerning the operation of the unit or need service refer to section 11 0 REQUESTING SERVICE When requesting service you will need the model and serial number as well as the purc...

Страница 3: ...Principles of Purifying Water 4 0 Product Specifications 5 0 Before Using Your Dispenser 5 1 Setup 5 2 Installing the Cartridges 5 3 Startup Procedure 5 4 Initial Polishing of new cartridges 6 0 Displ...

Страница 4: ...er cord grip the plug and pull it from the wall outlet NEVER PULL ON CORD DO NOT use this appliance outdoors This appliance is intended for commercial industrial institutional and professional use Use...

Страница 5: ...ure successful water system operation Trace organic chemicals can interfere with highly sensitive reactions and alter biological processes TDS Total Dissolved Solids conduct electricity This conductiv...

Страница 6: ...t a glance The quick connect cartridge system makes replacing filters a simple task For optimum performance the Aries High Purity Water System may require reverse osmosis distillation or deionizing pr...

Страница 7: ...sformer UV Combination for Bacteria and TOC Wall Mounting Kit Reverse Osmosis Pre treatment Packages Re circulating Spray Gun 0 05 micron Pyrogen removal filter Typical Applications Media Prep Tissue...

Страница 8: ...tion valve at this time 5 2 Installing the Cartridges There are three 3 cartridge positions The water flows from left to right Any pretreatment filters must be installed in position 1 furthest from th...

Страница 9: ...g to confirm the effectiveness of the treatment process is the owners responsibility 6 0 DISPLAY AND KEYPAD FUNCTIONS Be sure you understand the function of each button before using them Function Keys...

Страница 10: ...e top cover by removing the phillips head screws that attach it Gently lift the top cover up and left and set it aside Unclasp the clamps that hold the stainless steel body of the UV light Gently lift...

Страница 11: ...rom the outlet Place a wad of paper towels around each plunger to absorb the small amount water that will come out of the filters plungers and or upper fitting Remove the three cartridges starting wit...

Страница 12: ...s 1 5 0 micron prefilter 1 GAC carbon filter and 1 1 0 micron post filter 9 5 Aries HPL RO Replacement Membrane AM 12 7010 Replacement membrane is a thin film composite PA membrane with 100 GPD output...

Страница 13: ...se barb connection 9 12 Blank Sanitization Cartridges VP 17 0000 Blank VP Style cartridges for sanitization of unit 3 required per Aries System Sanitization kit sold separately 10 0 REQUESTING SERVICE...

Страница 14: ...turned on Bad Solenoid Have unit serviced by an authorized dealer or return to factory A filter cartridge is loaded with particles and plugged No flow Replace the filter cartridge Capsule filter is bo...

Страница 15: ...llege of American Pathologists ASTM American Society for Testing and Materials NCCLS National Committee for Clinical Laboratory Standards EU Endotoxin Units N L No Limit The measurements of pH in Type...

Страница 16: ...anty ARIES ResinTech Inc assumes no other responsibility or liability This warranty does not cover any product which has been subject to misuse neglect accident or used in violation of operating instr...

Отзывы: