24 Ethernet cable tests
ARGUS 165
289
Example measurements:
Open end
In this example, the line is open on all
pairs. ARGUS locates the fault in the area
from 49.3 m to 51.7 m.
This indicates that the line is open at
approx. 50 m.
Short
In this example, the line is short-circuited
between pairs 1-2 and 7-8. ARGUS
displays the value "Short" as the status.
Mismatch
This example shows a mismatch (>115
at pairs 3-6 and 4-5. This indicates an
open cable end or an impedance
transition >115