5.1 Functions which require pressing only 1 key: 

a / Display of the min. value 

b/ Display of the max. value

InF.

02.00

Display of 

the measure

Alternating infor-

mation display
Minimum value

5. FUNCTIONS WHICH CAN BE REACHED ON 

THE FRONT PANEL

SUP.

08.00

Display of 

the measure

Alternating infor-

mation display
Maximum value

c / Deleting of the min. and max. values

M

CLr.M

Display of 

the measure

The instrument will revert to the measure display.

Reminder:

 

If no key is pressed during 20 s., the

instrument will revert to the measure display.

Note:

 

An exit from mode programming with saving of the    

configuration will reset the min. and max. values to zero.

deleting of the recorded min. and 

max., and revert to the measure 

display

5.2 Functions which require pressing several keys

:

5.2.1 Shifting of the display

Shifting of the display down scale (AdJ.Lo)
Shifting of the display full scale (AdJ.Hi)

After injecting an input signal corresponding to the down (or full) display 

scale, press the keys 

and 

(or      and     ) simultaneously. The 

message AdJ.Lo (AdJ.Hi) will appear alternating with the value, to indi-

cate that you are in the menu adjustment.

By pressing       and       you can increase or decrease the down (or full) 

display scale.

If you keep pressing during 3s. on key      or      you can access a quick 

increasing or decreasing of the display value.

Press      to validae the shifting. The message OK will appear during the 

acknowledgement of the shifting (1s.), and the instrument will revert to 

mode measure. Once all the shifting are validated, the input thus shifted 

will keep this shifting even after a setting off tension. 

Press       (or do not press any key during 20 s) to revert to the measure 

display without modifications. 

• 

Case of a process, resistance or potentiometer input

The instrument will then re-adjust its scale factor and its display factor in 

order to obtain the required result on the display.

• 

Case of a temperature input

On a temperature input; if 1 of the 2 settings is performed: this will cor-

respond to an offset, which means that all the points will be shifted by 

the same quantity.
On the contrary, if the 2 settings are performed, the slope and the offset 

will be corrected in order to obtain the required result.

example

 : 

On a Pt100 input for 0°C, the obtained display is -000.3

For 500°C the obtained display is 0500.2. To correct this display, shift 

the display down scale by 3 points to obtain 000.0, and the display full 

scale by -2 points to obtain 0500.0.

M

M

M

(accessible according to the programmed 

access code)

Содержание DIP 10

Страница 1: ...ax 04 72 31 31 31 Tel Intern 33 4 72 31 31 30 Fax Intern 33 4 72 31 31 31 ARDETEM TA IN 46 C 11 06 Any data in this documentation may be modified without prior notice ARDETEM Soucieu en Jarrest FRANCE...

Страница 2: ...12 4 6 3 Safeties p13 4 6 4 Display features p14 4 7 Reading of the configuration p14 4 8 Access code p15 4 9 Programming of a new access code p15 4 10 Functions of the the main menu p15 4 10 1 Displa...

Страница 3: ...V typical 60 V Max CJC efficiency 0 03 C C 0 5 C from 5 C to 55 C MR Measure range Cut off the display and the output remain at down scale for an input signal value of the cut off programmable from 0...

Страница 4: ...Voltage direct or reversed 0 10V Load impedance Lr 500k 2 setpoints per relay configurable on the whole MR Hysteresis programmable from 0 to 100 Time delay programmable from 0 to 25 sec 8A 250 VAC on...

Страница 5: ...transmitter VOUT GND VCC 5 6 7 Tc Pt100 3 wire Ni 100 3 wire 4 5 6 7 Potentiometer RESISTANCE AND POTENTIOMETER TEMPERATURE PROCESS Resistance 440 up to 2k 0 8k optional Strap 4 and 7 on terminal 0 4...

Страница 6: ...ation display Entering of a parameter First start by increasing or decreasing the 1st digit and the sign from 9 to 9 the 2nd from 0 to 9 the 3rd from 0 to 9 the 4th from 0 to 9 M M Alarms Led 1 Led 2...

Страница 7: ...is and or sensor rupture and access to disconnecting the sensor rupture option analog or relay output Access to the programming of the display p10 Access to the programming exit menu with p10 or witho...

Страница 8: ...d the analog output SPxx hystx do diS Fo diS If thermocouple if other than thermocouple CJC E CJ t CJC yES no AdJuS to tc diSPL tc S tc CA tc J tc r tc t tc E tc b tc n tc C tc C3 tc C5 tc L oF oC rES...

Страница 9: ...yES 88 88 1 1 setting in display points You can choose to assign the analog output to the input measure PArA InPut or to the displayed measure PArA diSPL Eg with PArA diSPL When the transmitter indic...

Страница 10: ...tus of the relay 1 in case of sensor rupture LO HI OFF status of the relay 2 in case of sensor rupture disconnection of the sensor rupt or yES no value of the output in case of sensor rupture 00 00 rE...

Страница 11: ...max as well as the alarm recordings 4 5 1 Current input Special linearisation For specific applications such as the measurement of volume the meter can memorise an unlinear curve programmable in X an...

Страница 12: ...A F in 20 mA nb 9 d disp 0 000 m3 F disp 0 785 m3 Programming from A01 to A09 and from B01 to B09 according to the table 0 2 4 6 8 10 12 14 16 18 20 0 0 1 0 2 0 3 0 4 0 5 0 6 0 7 0 8 0 9 1 20 18 16 14...

Страница 13: ...m as the led represents the alarm state OFF OFF OFF ON ON ON setpoints setpt setpt or SP X SP X HYSx HySt x LEdx On OFF SP1 X SP1 X HYSt x SP2 X SP2 X HYSt x SP2 X SP1 X SP2 X SP1 X led lit led lit le...

Страница 14: ...t supplied in case of sensor rupture Relay active coil supplied in case of sensor rupture Note the led is either still or lit according to its programming in the menu rELAY 4 6 3 Safeties Self diagnos...

Страница 15: ...ration indice of the digital filtering Programmable from 0 to 10 for use in case of unsteady input signal To obtain the maximum response time add 240 ms Note For the response time of the analog output...

Страница 16: ...XXXXX 1 2 3 4 5 P CodE OLd no Enter former code if code not valid old display during 2s and revert to the measure display New code registered and revert to the measure display if code correct old Ente...

Страница 17: ...orresponding led starts to blink To revert to the normal status led not blinking and relay in the correct state use menu CLEAr M CLEAr the recorded alarms are not deleted deleting of the recorded alar...

Страница 18: ...can increase or decrease the down or full display scale If you keep pressing during 3s on key or you can access a quick increasing or decreasing of the display value Press to validae the shifting The...

Страница 19: ...Setting of the setpoints There are 2 ways to adjust setpoints either in mode programming entering the correct safety access code see p15 or by pressing simultaneously on and The meter will then show...

Страница 20: ...UE Messages shown by the meter in mode programming and or in mode reading General access Access to the reading of the parameters Access to the programming of the input and output parameters Code for a...

Страница 21: ...l point in mode measure Place of the decimal point Choice of the processing function Linear Input down scale Input full scale Down and full display scale diSPL Point Funct LinEA d in F in F diSP d diS...

Страница 22: ...oints Sepoint and hysteresis in display points Operating mode of the relay x Mode setpoint Mode window Value of the setpoint in mode setpoint Value of the 1st setpoint in mode window Value of the 2nd...

Страница 23: ...ing of the instrument internal features Access to the submenu reading of the internal features Name of the instrument Type of the instrument Available options standard A analog output R relay output I...

Отзывы: