5.1 Functions which require pressing only 1 key:
a / Display of the min. value
b/ Display of the max. value
InF.
02.00
Display of
the measure
Alternating infor-
mation display
Minimum value
5. FUNCTIONS WHICH CAN BE REACHED ON
THE FRONT PANEL
SUP.
08.00
Display of
the measure
Alternating infor-
mation display
Maximum value
c / Deleting of the min. and max. values
M
CLr.M
Display of
the measure
The instrument will revert to the measure display.
Reminder:
If no key is pressed during 20 s., the
instrument will revert to the measure display.
Note:
An exit from mode programming with saving of the
configuration will reset the min. and max. values to zero.
deleting of the recorded min. and
max., and revert to the measure
display
5.2 Functions which require pressing several keys
:
5.2.1 Shifting of the display
Shifting of the display down scale (AdJ.Lo)
Shifting of the display full scale (AdJ.Hi)
After injecting an input signal corresponding to the down (or full) display
scale, press the keys
and
(or and ) simultaneously. The
message AdJ.Lo (AdJ.Hi) will appear alternating with the value, to indi-
cate that you are in the menu adjustment.
By pressing and you can increase or decrease the down (or full)
display scale.
If you keep pressing during 3s. on key or you can access a quick
increasing or decreasing of the display value.
Press to validae the shifting. The message OK will appear during the
acknowledgement of the shifting (1s.), and the instrument will revert to
mode measure. Once all the shifting are validated, the input thus shifted
will keep this shifting even after a setting off tension.
Press (or do not press any key during 20 s) to revert to the measure
display without modifications.
•
Case of a process, resistance or potentiometer input
The instrument will then re-adjust its scale factor and its display factor in
order to obtain the required result on the display.
•
Case of a temperature input
On a temperature input; if 1 of the 2 settings is performed: this will cor-
respond to an offset, which means that all the points will be shifted by
the same quantity.
On the contrary, if the 2 settings are performed, the slope and the offset
will be corrected in order to obtain the required result.
example
:
On a Pt100 input for 0°C, the obtained display is -000.3
For 500°C the obtained display is 0500.2. To correct this display, shift
the display down scale by 3 points to obtain 000.0, and the display full
scale by -2 points to obtain 0500.0.
M
M
M
(accessible according to the programmed
access code)