Arctiko UPUL 580 Скачать руководство пользователя страница 3

IMPORTANT
Before using the freezer

For safety reasons, the instructions should be read before the freezer is commissioned. 
The operating instructions should always be available and accessible to personnel!

The appliance concerned complies with the following directives and standards except for refrigerant 

amount:

Low Voltage Directive 2006/95/EEC

Electromagnetic Compatibility Directive 2004/108/EEC
Only use the freezer for its specific purpose.

Safeguarding the environment:
1. Packing

The packing material is entirely recyclable and marked with a recycling symbol.
Please follow the local regulations for scrapping.

2. Appliance

The freezer is made from recyclable material. Disposal of the appliance should be carried out according to 

local environmental regulations.

Make the appliance unusable by cutting of the power supply cord.  Please deliver the appliance to an 

authorised centre for recovery of refrigerant.  

Information

This appliance contains a mixture of various refrigerants. For more information please refer to the rating 

plate affixed to the freezer.

Precautions and general recommendations:

WARNING:

Keep ventilation openings, in the appliance enclosure or in the built-in structure, clear of 

obstruction.

WARNING:

Do not use mechanical devices or other means to accelerate the defrosting process, other than 

those recommended by the manufacturer.

WARNING:

Do not damage the refrigerant system.

WARNING:

Do not use electrical appliances inside the freezer storage compartment, unless they are of the 

type recommended by the manufacturer.

Always keep the keys in separate place and out of reach of children.

Before servicing or cleaning the appliance, unplug it from the mains or disconnect the electrical 

power supply.

The power supply cable may only be replaced by an authorised person.

Frost formation on the interior wall and upper parts is a natural phenomenon, and therefore the 

freezer should be defrosted during normal cleaning or maintenance. 

3

Содержание UPUL 580

Страница 1: ...OPERATING INSTRUCTIONS UPUL 580...

Страница 2: ...scription of the control panel 6 Access code 6 Turn Push button 7 Main switch 7 Menu selection 7 Adjustment of settings 7 Operation description 7 Usage 8 Setup 8 Manual mode 8 Calibration of sensor 8...

Страница 3: ...ed centre for recovery of refrigerant Information This appliance contains a mixture of various refrigerants For more information please refer to the rating plate affixed to the freezer Precautions and...

Страница 4: ...rities contact professional instructors or repair technicians immediately Installing the Freezer 1 You will find 3 different plugs for the installation Use the plug suitable for your electrical connec...

Страница 5: ...pliance be earthed The manufacture declines all liability for injury to persons or animals and for damage to property resulting from failure to observe the above procedures and reminders If the socket...

Страница 6: ...not to effect the inside temperature of the unit Disconnecting the appliance Electrical disconnecting must be possible either by unplugging the appliance or by means of a double pole switch located up...

Страница 7: ...until the required parameter value is shown in the display Press once to edit the value It will now flash Turn the button to adjust the parameter Press the button to accept the value or cancel it by...

Страница 8: ...eezer temperature rises above SetPoint TDiff they will start up again Please note that the fan will run 15 seconds before and after the compressor Manual mode A person with the correct access code wil...

Страница 9: ...he Alarm Menu can be activated Delete an inactive alarm from the list by pushing the Turn Push Button Active alarms cannot be acknowledged In the ALARM LOG sub menu the last 16 alarms can be read If t...

Страница 10: ...n the set temperature and the temperature shown will be the same CO2 Back Up 2 4 x port holes are prepared for CO back up and an injection tube see separate enclosure can be inserted 2 from the top of...

Страница 11: ...is not recommended to lay down the unit Please let the unit rest in vertical position for minimum 12 hours before start up Cleaning and Maintenance of the Freezer Ice will form on the inner doors of t...

Страница 12: ...e unit type number and the function of the part s concerned Disposing of the Freezer To prevent harmful environmental effects contact the distributor and ask where the freezer should be delivered Remo...

Страница 13: ...button for 10 seconds Can also be done by the user This menu is applied by the technician to decide whether the freezer should run automatically or manually View menu shows set point and the number of...

Страница 14: ...this menu the technician can programme the freezer to run automatically or manually Back to previous menu Shows current set point Shows how many alarms are saved in the alarm log Back to previous menu...

Страница 15: ...ing of the lowest alarm limit TA1Lo TSet Alarm The programming of the highest alarm limit TA1Hi TSet Alarm The delay of the abovementioned alarms starts The technician has to programme the maximum all...

Страница 16: ...with reference to the alarm Shows the last 15 alarms Back to previous menu Back to previous menu Shows which software version is installed in the unit Shows the total running hours of compressor Shows...

Страница 17: ...of Fan Manual running of light Manual running of CTS Test of Power Sence Relay When this relay runs out it is discovered by D12 Test of battery and test of Battery Disconnect Relay Manual running of...

Страница 18: ...18 CONNECTION DIAGRAM...

Страница 19: ...19...

Страница 20: ...LAMMEFJORDSVEJ 5 DK 6715 ESBJERG N DENMARK TEL 45 70 20 03 28 FAX 45 70 20 03 29 INFO ARCTIKO COM WWW ARCTIKO COM WWW ARCTIKO COM 50 80 014 xx GB...

Отзывы: