Arctiko UPUL 580 Скачать руководство пользователя страница 11

11

Filling procedure:

Empty the system and evacuate to 2 mbar. Evacuation must be performed at two places in the 

system; both from the filling pipe and from the pipe mounted on the pressure side of the compressor.

Mount the capillary pipe with mounting screws between the filling station and the service bottle.

Unlike with other refrigerants, the service bottle must be turned tap-down.

Commence filling.

The compressor will start after approx. 10-15 seconds of filling.

The bottle is empty when the pressure falls to 1bar and must be shut immediately. Immediate 

shutting of the bottle is very important to prevent the fluid being pushed back into the bottle.

Close the system.

Relocating or Moving the Freezer

Never leave the freezer to others without instructing those people in its use and the safety regulations 

described in the operating instructions. Always let an authorised electrician install the freezer. (Remember to 

wait for one hour before restarting it).

If the freezer and its electrical parts are reconstructed, the guarantee and safety regulations shall cease to 

apply.

If the freezer does not work: call an authorised servicing engineer!

Contact the distributor or an authorised service repair shop!

It is not recommended to lay down the unit. Please let the unit rest in vertical position for minimum 12 

hours before start up.

Cleaning and Maintenance of the Freezer:

Ice will form on the inner doors of the freezer. This ice should be removed with a scraper at regular 

intervals. (Remember safety gloves).

Ice on the inside of the cabinet should be removed as required. Take the contents out of the freezer. 

Pull out the plug and leave the door open until the ice can be removed. (Remember safety gloves).

Dry the inner cabinet and shelves with a cloth until it is not damp any longer.

For external maintenance use stainless steel polish.

Ice formed on the inside of the door should be removed with a wooden/plastic scraper. This ice will 

build up because of the low temperatures inside the cabinet, and if it is not removed it will cause the 

door to open.

Cleansing agents containing chlorine or compounds of chlorine as well as other corrosive means 

may not be used as they might corrode the stainless panels of the cabinet and the evaporator 

system.

The compressor compartment and in particular the condenser must be kept free from dust and dirt. 

This is best done with a vacuum cleaner and a brush.

Do not put bottles containing carbonic acid or inflammable products in the freezer.
Disposal and scrapping of the freezer: Remove the lock, so that there is no risk of children getting locked in 

the freezer whilst playing. 

Содержание UPUL 580

Страница 1: ...OPERATING INSTRUCTIONS UPUL 580...

Страница 2: ...scription of the control panel 6 Access code 6 Turn Push button 7 Main switch 7 Menu selection 7 Adjustment of settings 7 Operation description 7 Usage 8 Setup 8 Manual mode 8 Calibration of sensor 8...

Страница 3: ...ed centre for recovery of refrigerant Information This appliance contains a mixture of various refrigerants For more information please refer to the rating plate affixed to the freezer Precautions and...

Страница 4: ...rities contact professional instructors or repair technicians immediately Installing the Freezer 1 You will find 3 different plugs for the installation Use the plug suitable for your electrical connec...

Страница 5: ...pliance be earthed The manufacture declines all liability for injury to persons or animals and for damage to property resulting from failure to observe the above procedures and reminders If the socket...

Страница 6: ...not to effect the inside temperature of the unit Disconnecting the appliance Electrical disconnecting must be possible either by unplugging the appliance or by means of a double pole switch located up...

Страница 7: ...until the required parameter value is shown in the display Press once to edit the value It will now flash Turn the button to adjust the parameter Press the button to accept the value or cancel it by...

Страница 8: ...eezer temperature rises above SetPoint TDiff they will start up again Please note that the fan will run 15 seconds before and after the compressor Manual mode A person with the correct access code wil...

Страница 9: ...he Alarm Menu can be activated Delete an inactive alarm from the list by pushing the Turn Push Button Active alarms cannot be acknowledged In the ALARM LOG sub menu the last 16 alarms can be read If t...

Страница 10: ...n the set temperature and the temperature shown will be the same CO2 Back Up 2 4 x port holes are prepared for CO back up and an injection tube see separate enclosure can be inserted 2 from the top of...

Страница 11: ...is not recommended to lay down the unit Please let the unit rest in vertical position for minimum 12 hours before start up Cleaning and Maintenance of the Freezer Ice will form on the inner doors of t...

Страница 12: ...e unit type number and the function of the part s concerned Disposing of the Freezer To prevent harmful environmental effects contact the distributor and ask where the freezer should be delivered Remo...

Страница 13: ...button for 10 seconds Can also be done by the user This menu is applied by the technician to decide whether the freezer should run automatically or manually View menu shows set point and the number of...

Страница 14: ...this menu the technician can programme the freezer to run automatically or manually Back to previous menu Shows current set point Shows how many alarms are saved in the alarm log Back to previous menu...

Страница 15: ...ing of the lowest alarm limit TA1Lo TSet Alarm The programming of the highest alarm limit TA1Hi TSet Alarm The delay of the abovementioned alarms starts The technician has to programme the maximum all...

Страница 16: ...with reference to the alarm Shows the last 15 alarms Back to previous menu Back to previous menu Shows which software version is installed in the unit Shows the total running hours of compressor Shows...

Страница 17: ...of Fan Manual running of light Manual running of CTS Test of Power Sence Relay When this relay runs out it is discovered by D12 Test of battery and test of Battery Disconnect Relay Manual running of...

Страница 18: ...18 CONNECTION DIAGRAM...

Страница 19: ...19...

Страница 20: ...LAMMEFJORDSVEJ 5 DK 6715 ESBJERG N DENMARK TEL 45 70 20 03 28 FAX 45 70 20 03 29 INFO ARCTIKO COM WWW ARCTIKO COM WWW ARCTIKO COM 50 80 014 xx GB...