Only qualified/authorized engineers or service personnel should 

install the appliance

. Installation by unqualified personnel may cause 

electric shock or fire.

Always make sure the appliance is grounded to protect the user from 

electrical shock.

Do not use the power supply cord if it is damagaged.

 Such supply 

cord may cause fire or electric.

Never use a telephone line or lightning rod as ground protection. 

During lightening, there is a strong current present, which is 

extremely dangerous.

Do not use water pipes as ground protection. Modern water pipe 

systems are non-conductors such as PVC.

Replacing the power cord may only be done by authorized 

personal.

!

Never use gas lines as the ground protection for the appliance. This 

can cause an explosion.

Disconnect the power cord if there is something wrong with the 

unit. 

Continued abnormal operation may cause electric shock or fire.

Electrical connection

In order to ensure a reliable installation that complies with the limits of 

temperature and voltage drop, it is necessary to determine maximum load of 

appliances connected to the installation. In determining the maximum load for 

an installation or for a part of it, it is vital to take into account 

contemporaneous factors. For supply systems, the following must be 

determined:

- Power system (AC/DC) - Data regarding voltage and absorbed power/

  current is given on the rating plate. 

- Ground protection -

 

To prevent the user of getting exposed to electrical 

  shock, in the event of a insulation damage, the appliance must be grounded. 

- The installation must always be secured with a minimum 10 A fuse.

- If more than one appliance needs to be installed, each appliance  

  must be connected to an individual fuse group. 

 

When installing the appliance, make sure the protected earth is grounded. If 

the connection is a 3-prong connection, use a three-pin plug and connect the 

conductor with yellow / green insulation to ground. In order to maintain a 

stable operation of the appliance, voltage variation cannot differ more than 

±10 percent of the nominal voltage supply.  

Always follow local regulations when preparing an installation. If in doubt 

always contact your local authorities.

Getting started

During the initial startup and continuous usage of the appliance, the following 

procedures shall be followed.

1.  Plug the power socket into a dedicated outlet. For correct voltage 

requirement, follow the information on the rating plate.

When started, let the appliance cool for 24 hours before placing any 

products into it.

 

2.  When the appliance initially starts up, the high temp. alarm may be 

activated. This is normal as the chamber temperature may be outside the 

alarm limits during the temperature pull down. You can safely mute or 

cancel the alarm (see section “How to mute and accept alarms”).

3.  This appliance has been set to operate at -80°C (ultra low temp. freezers) 

/ -40°C (low temp. freezers) from the factory.

4.  Check the functionality of the light by opening the door. 

5.  Once a thorough inspection of the appliance is completed, products can 

be loaded into the appliance. Products should be pre-frozen when inserted 

into the appliance, otherwise this can affect the cooling performance of 

the appliance.

UNIT

AMPS (A)

ULUF 15

4

ULUF 65

5

ULUF 125

5

ULUF 450

7

ULUF 550

9

ULUF 750

12

Maximum current rating (ambient 20°C)

Содержание ULUF

Страница 1: ...ORIGINAL OPERATING INSTRUCTIONS ULUF models Ultra low temperature freezers 90 C Ultra low temperature freezers 86 C Low temperature freezers 40 C...

Страница 2: ...her information about the appliance Symbols Abbreviations Safety Intended use Personal safety Handling Transportation Moving the appliance Packaging Unpacking Inactivity for extended period Disposal I...

Страница 3: ...the advantages of your appliance To meet the high quality demands required by the medical and pharmaceutical industry at least 10 of all appliances leaving Arctiko are carefully checked and tested to...

Страница 4: ...te the appliance and make sure it is completely dry Disconnect the power cord Leave the door slightly open in order to prevent rot and mold Disposal In the event of disposal of the appliance observe r...

Страница 5: ...ut the refrigerant Location Place the appliance in a location that satisfies the following conditions in or der to achieve optimal operating results Firm and levelled floor Placing the appliance on a...

Страница 6: ...on must always be secured with a minimum 10 A fuse If more than one appliance needs to be installed each appliance must be connected to an individual fuse group When installing the appliance make sure...

Страница 7: ...w and low temperature freezing with high performance and stability Standard features is filter less construction for less maintenance low noise level low energy consumption low heat dissipation and ex...

Страница 8: ...ion Never insert probes through the door as this will danger the door gasket increase the ice build up and affect the temperature inside the appliance Display G214 The control panel is a multifunction...

Страница 9: ...lay is mounted in the top panel 17 Push the button in order to access a temperature graph Push the buttons to navigate the data history for up to 10 days Push the Esc button in order to exit the menu...

Страница 10: ...as a backup guaranteeing that your stored products will be kept at the set temperature An Arctiko appliance which offers dual cooling system can be identified by the denomination of D in the end of t...

Страница 11: ...set between the minimum and maximum value 21 Controller operation settings Custom settings Advanced Settings Advanced service settings Status Change Reset password 20 How to change the setpoint Alarm...

Страница 12: ...imit for a period longer than the Alarm delay the low temperature alarm is activated If products with a lower temperature than the low temperature alarm setting is placed against or close to the tempe...

Страница 13: ...r will not be reactivated by accepting the alarm Recommended actions If it is a non recurring alarm the compressor can be reactivated by restarting the appliance switching the appliance off and on aga...

Страница 14: ...27 Select language The default language of the controller is English The language can be changed in the Select language menu The available languages are English German Russian French Spanish Swedish...

Страница 15: ...optional extra probes Probe 2 and Probe 3 for temperature monitoring The chamber probe is measuring the air temperature inside the appliance to control the operation of the cooling system when the co...

Страница 16: ...fect the performance of the appliance negatively or cause damage to the appliance The technical data can be found on Arc tiko s website or by contacting the seller of the appliance 31 Service The serv...

Страница 17: ...ess Hysteresis Press ENTER to access Temp range limits ATTENTION Do not change the hysteresis setting without contacting a service representative Use to choose probe Press ENTER to access Use to chang...

Страница 18: ...ory setting 0000 In order to reset the password you need to require a reset password from your Arctiko distributor Contact your distributor for the code needed Press ENTER to access menu Press until C...

Страница 19: ...or similar software This function also makes it possible for a service technician to analyze the data for troubleshooting in case of malfunctions for quick and accurate support Insert USB drive in th...

Страница 20: ...eadout has been preformed each readout will create a set of files with the number increased by one The data00 txt file contains the most recent download The param00 txt file contains the parameter set...

Страница 21: ...d Compressor alarm Marked with H if triggered Manual defrost Marked with H if cycle is running Auto defrost Marked with H if cycle is running Measurements tmp1 Temperature measured by 1st probe tmp2 T...

Страница 22: ...rost Freezer 0 Manual defrost Freezer OK Hysteresis Freezer 2 0 Temp range limits Freezer Max 20 0 Freezer Min 40 0 42 Default settings ULUF 15 65 125 Description Arctiko default settings Customer set...

Страница 23: ...0 Manual defrost Freezer OK Hysteresis Freezer 2 0 Temp range limits Freezer Max 40 0 Freezer Min 85 0 44 Default settings ULUF 490 490 2M 590 590 2M 890 890 2M Description Arctiko default settings Cu...

Страница 24: ...mends a thorough examination twice a year and cleaning at least once a month General maintenance Perform following at least twice a year Lubricate hinges and gaskets Wipe off all excess lubricant Make...

Страница 25: ...ter can damage the insulation materials and electrical components Dust off the inside of the compressor compartment using only a dry brush or a vacuum cleaner Parts of the refrigeration system of this...

Страница 26: ...m the sunlight or heat source Frequent door openings may cause the temperature to rise Check if there has been frequent door openings Leave the door closed until the temperature in the appliance is st...