Safety

Arctiko is obligated as a supplier to ensure the users safety when operating 

one of our appliances. To prevent personal and material injury or damage, 

please follow the instructions in this manual.

Intended use 

Ultra low temperature and low temperature appliances are designed for 

storage of biomedical products at a constant temperature in clinical, 

pharmaceutical, research and laboratory fields. The appliances are designed 

to operate within a temperature range of  -90°C to -40°C, -86°C to -40°C or 

-40°C to -20°C depending on the model and at a maximum ambient 

temperature of +25°C. 

For safe and optimal performance of the appliance, it should only be placed 

indoors, in a well ventilated room and in elevations below 2000 m. The 

appliance should only be operated by instructed personnel. 

Personal safety 

When operating the appliance, your body is exposed to a variation of possible 

dangerous events. 

- Read and understand this manual. If in doubt, contact your local Arctiko   

  distributor.

- Use freezer gloves at all times when loading or unloading the appliance. 

  The operating temperature is such that direct contact with the cold 

  contents or inside the appliance can burn unprotected skin. 

- Assure good cleaning practices at all times by keeping the appliance and 

  the adjacent areas clean, dry and uncluttered.

Introduction

!

Abbreviations

Symbols

Please take a few minutes to read and become familiar with the advantages 

of your appliance. To meet the high quality demands required by the medical 

and pharmaceutical industry, at least 10 % of all appliances leaving Arctiko are 

carefully checked and tested to ensure high performance and quality. If the op-

erating and installation instructions describe different models, any differences 

will be pointed out at the relevant points in the text. Read these operating and 

installation instructions before switching on the appliance.

If you would like to obtain further information about this appliance, please visit 

our webside www.arctiko.com where downloads are available on the page of 

each specific product.

4

5

!

!

!

A       

Amps

V       

Volt

     Watt

C       

Degree Celcius

Kg     

Kilogram

MM 

  Milimeter

Min

     Minutes

H

         Hour

N/A

     Not applicable

ULT

     Ultra-low temperature

WARNING 

Performing this action can cause personal injury.

CAUTION 

Risk of personal or material injury

Consult the instructions before attempting to use this equipment.

Prohibition  

Action is strictly prohibited.

Follow procedures  

Keep the instructions handy for convenient reference.

OFF

Disconnect power supply before operation.

Fragile

 -

 Handle with with care

Indicates a device that can be damaged or broken if not 

handled with care.

Grounding

Be aware that the appliance is grounded.

Do not insert metal objects such as pins or a wires into any vent, 

gap or any outlet on the appliance. 

This may cause electric shock or 

injury by accidental contact with moving parts.

When removing the power plug from the outlet, pull on the plug, 

not the cord. 

Pulling the power cord may result in electric shock or 

fire by short circuit.

Содержание ULUF

Страница 1: ...ORIGINAL OPERATING INSTRUCTIONS ULUF models Ultra low temperature freezers 90 C Ultra low temperature freezers 86 C Low temperature freezers 40 C...

Страница 2: ...her information about the appliance Symbols Abbreviations Safety Intended use Personal safety Handling Transportation Moving the appliance Packaging Unpacking Inactivity for extended period Disposal I...

Страница 3: ...the advantages of your appliance To meet the high quality demands required by the medical and pharmaceutical industry at least 10 of all appliances leaving Arctiko are carefully checked and tested to...

Страница 4: ...te the appliance and make sure it is completely dry Disconnect the power cord Leave the door slightly open in order to prevent rot and mold Disposal In the event of disposal of the appliance observe r...

Страница 5: ...ut the refrigerant Location Place the appliance in a location that satisfies the following conditions in or der to achieve optimal operating results Firm and levelled floor Placing the appliance on a...

Страница 6: ...on must always be secured with a minimum 10 A fuse If more than one appliance needs to be installed each appliance must be connected to an individual fuse group When installing the appliance make sure...

Страница 7: ...w and low temperature freezing with high performance and stability Standard features is filter less construction for less maintenance low noise level low energy consumption low heat dissipation and ex...

Страница 8: ...ion Never insert probes through the door as this will danger the door gasket increase the ice build up and affect the temperature inside the appliance Display G214 The control panel is a multifunction...

Страница 9: ...lay is mounted in the top panel 17 Push the button in order to access a temperature graph Push the buttons to navigate the data history for up to 10 days Push the Esc button in order to exit the menu...

Страница 10: ...as a backup guaranteeing that your stored products will be kept at the set temperature An Arctiko appliance which offers dual cooling system can be identified by the denomination of D in the end of t...

Страница 11: ...set between the minimum and maximum value 21 Controller operation settings Custom settings Advanced Settings Advanced service settings Status Change Reset password 20 How to change the setpoint Alarm...

Страница 12: ...imit for a period longer than the Alarm delay the low temperature alarm is activated If products with a lower temperature than the low temperature alarm setting is placed against or close to the tempe...

Страница 13: ...r will not be reactivated by accepting the alarm Recommended actions If it is a non recurring alarm the compressor can be reactivated by restarting the appliance switching the appliance off and on aga...

Страница 14: ...27 Select language The default language of the controller is English The language can be changed in the Select language menu The available languages are English German Russian French Spanish Swedish...

Страница 15: ...optional extra probes Probe 2 and Probe 3 for temperature monitoring The chamber probe is measuring the air temperature inside the appliance to control the operation of the cooling system when the co...

Страница 16: ...fect the performance of the appliance negatively or cause damage to the appliance The technical data can be found on Arc tiko s website or by contacting the seller of the appliance 31 Service The serv...

Страница 17: ...ess Hysteresis Press ENTER to access Temp range limits ATTENTION Do not change the hysteresis setting without contacting a service representative Use to choose probe Press ENTER to access Use to chang...

Страница 18: ...ory setting 0000 In order to reset the password you need to require a reset password from your Arctiko distributor Contact your distributor for the code needed Press ENTER to access menu Press until C...

Страница 19: ...or similar software This function also makes it possible for a service technician to analyze the data for troubleshooting in case of malfunctions for quick and accurate support Insert USB drive in th...

Страница 20: ...eadout has been preformed each readout will create a set of files with the number increased by one The data00 txt file contains the most recent download The param00 txt file contains the parameter set...

Страница 21: ...d Compressor alarm Marked with H if triggered Manual defrost Marked with H if cycle is running Auto defrost Marked with H if cycle is running Measurements tmp1 Temperature measured by 1st probe tmp2 T...

Страница 22: ...rost Freezer 0 Manual defrost Freezer OK Hysteresis Freezer 2 0 Temp range limits Freezer Max 20 0 Freezer Min 40 0 42 Default settings ULUF 15 65 125 Description Arctiko default settings Customer set...

Страница 23: ...0 Manual defrost Freezer OK Hysteresis Freezer 2 0 Temp range limits Freezer Max 40 0 Freezer Min 85 0 44 Default settings ULUF 490 490 2M 590 590 2M 890 890 2M Description Arctiko default settings Cu...

Страница 24: ...mends a thorough examination twice a year and cleaning at least once a month General maintenance Perform following at least twice a year Lubricate hinges and gaskets Wipe off all excess lubricant Make...

Страница 25: ...ter can damage the insulation materials and electrical components Dust off the inside of the compressor compartment using only a dry brush or a vacuum cleaner Parts of the refrigeration system of this...

Страница 26: ...m the sunlight or heat source Frequent door openings may cause the temperature to rise Check if there has been frequent door openings Leave the door closed until the temperature in the appliance is st...

Страница 27: ...5080038 11 GB...