30
Hysteresis
The hysteresis in combination with the setpoint defines the temperatures for
starting and stopping the cooling compressor in the appliance. E.g. with a
setpoint of -80°C and a hysteresis of 2, the compressor will start when the
temperature in the appliance reaches -79°C and stop again at -81°C.
Attention!
Do not change the hysteresis setting without contacting a service
representative. If set incorrectly, it may severely affect the stability and
cooling performance of the appliance.
Temp. range limits
The temperature range limits define the highest and lowest temperature the
setpoint of the appliance can be set to.
Attention!
Changing these parameters and operating the appliance outside of the
specifications listed in the technical data may affect the performance of the appliance
negatively or cause damage to the appliance. The technical data can be found on Arc-
tiko’s website or by contacting the seller of the appliance.
31
Service
The service menu offers information on compressor and fan activation time,
probe temperatures and firmware versions of controller. There are no
changeable settings in the service menu.
Password protection ON/OFF device
When turning ON/OFF, the appliance password protection can be enabled
to avoid accidentally switching off the appliance. If activated, a password for
either the custom settings or advanced settings has to be entered to turn the
appliance on or off.
Set date/time
The date and time is displayed on the main screen and has to be manually
updated to match the local time.
Ambient temp. settings
If the ambient temperature shown in the display varies from the actual room
temperature, the shown temperature can be calibrated to match the actual
temperature.