Default settings ULUF 450/450-2M/550/550-2M/750/750-2M/850/850-2M:
Description
Arctiko default
settings
Customer
settings
Custom Settings
(Password)
0000
Set point
Freezer
-80,0
Alarm Settings:
Alarm delay
Freezer
15 min
Door open alarm
Freezer
Enabled
High temp. alarm
Freezer
-70
Low temp. alarm
Freezer
-90
Probe/eprom failure
Freezer
Enabled
Power failure
Freezer
Enabled
Logging time interval
Freezer
1
Description
Arctiko default
settings
Customer
settings
Advanced Settings (Password)
0000
Calibration
Freezer
0,0
Automatic defrost
Freezer
0
Manual defrost
Freezer
OK
Hysteresis
Freezer
2,0
Temp. range limits
Freezer Max.
-40,0
Freezer Min.
-85,0
44
Default settings ULUF 490/490-2M/590/590-2M/890/890-2M:
Description
Arctiko default
settings
Customer
settings
Custom Settings
(Password)
0000
Set point
Freezer
-80,0
Alarm Settings:
Alarm delay
Freezer
15 min
Door open alarm
Freezer
Enabled
High temp. alarm
Freezer
-70
Low temp. alarm
Freezer
-100
Probe/eprom failure
Freezer
Enabled
Power failure
Freezer
Enabled
Logging time interval
Freezer
1
Description
Arctiko default
settings
Customer
settings
Advanced Settings (Password)
0000
Calibration
Freezer
0,0
Automatic defrost
Freezer
0
Manual defrost
Freezer
OK
Hysteresis
Freezer
2,0
Temp. range limits
Freezer Max.
- 0,0
Freezer Min.
- 0,0
45