Model 150A400
Manual Text
2-4
REV B
2.2.2
Remote Operation
(continued)
2.2.2.4.2 IEEE-488 device address selection
(continued)
Table 2-1
IEEE-488 Device Address Selection
Device Address
Switch 5
Switch 4
Switch 3
Switch 2
Switch 1
1
off (0)
off (0)
off (0)
off (0)
on (1)
2
off (0)
off (0)
off (0)
on (1)
off (0)
3
off (0)
off (0)
off (0)
on (1)
on (1)
4
off (0)
off (0)
on (1)
off (0)
off (0)
5
off (0)
off (0)
on (1)
off (0)
on (1)
:
:
30
on (1)
on (1)
on (1)
on (1)
off (0)
2.2.2.5
Command Set Format
Each command is composed of one alpha character, up to four numeric parameters, and a
command termination character. The command termination character is the “line feed”
command, which is denoted and entered as <LF>. Commands are case-sensitive and must be
entered in upper case only in order to be recognized.
2.2.2.6
IEEE-488 Communications
For IEEE-488 communications, the “End or Identify” (EOI) control line may also be used for
command termination. When sending commands to the Model
150A400
via the IEEE-488 bus,
terminate each command with a <LF>, an EOI, or both. No characters are permitted after the
<LF> or EOI; the
150A400
interprets characters following the <LF> or EOI as the start of the
next command. When an error condition is present at the Model
150A400
, the
“Service Request” (SRQ) line is asserted; the operator can then perform a serial poll operation.
The Model
150A400
error code (in binary) is contained in the returned serial poll status byte
(STB). These error codes are defined in
Table 2-2
.