Model 150A400
Manual Text
2-3
REV B
2.2.2.2
Selecting Remote Operation
The Model 150A400 can be placed in the remote operation mode at any time by switching the
FUNCTION
switch on the front panel to the
REMOTE
position. In this mode, control is
transferred to the selected remote interface and all front panel controls are inoperative with the
exception of the
FUNCTION
switch. The amplifier’s initial state will be
Power Off
. The front
panel VFD will indicate
REMOTE
until the unit is returned to the local operation mode.
2.2.2.3
Interface Selection
The Model
150A400
can be controlled via either the IEEE-488 or RS-232 interface; which
interface is active is determined by the position of Switch 6 of the rear panel Dual In-Line
Package (DIP) switch located between the two interface connectors. If Switch 6 is in the “on”
(
1
) position, the RS-232 interface will be active; if Switch 6 is in the “off” (
0
) position, the
IEEE-488 interface will be active.
2.2.2.4
Interface Set-up
Switches 1–5 of the rear panel DIP switch are used to select either the RS-232 communication
(BAUD) rate or the IEEE-488 device address, depending upon which interface is active. (
Note:
These switches are only read at device power-up. In order for changes made in these switch
settings to take place, AC power must be removed and then re-applied to the Model
150A400
.)
2.2.2.4.1 RS-232 BAUD rate selection
The serial communication (BAUD) rate can be set to five different levels. Selections are made
by the positions of Switches 1–5 of the rear panel DIP switch. The following is a list of the
available BAUD rates and the corresponding DIP switch positions:
BAUD Rate
Switch On (1)
1200
1 only
2400
2 only
9600
3 only
19,600
4 only
76,800
5 only
(
Note:
Any other combination of switch settings will result in a BAUD rate equal to 1200.)
2.2.2.4.2 IEEE-488 device address selection
The IEEE-488 device address can be set to any number between 1 and 30. This selection is
made by setting Switches 1–5 of the rear panel DIP switch to the binary equivalent of the
number.
Table 2-1
illustrates this switch selection.