4-3
REV B
4.3.1.2
Check the
FUNCTION
switch on the unit’s front panel; it must be set to the
LOCAL
position in order to operate the front panel
POWER
switch. Check the circuit breaker on
the unit’s rear panel; it must be set to the “
1
” (“ON”) position.
4.3.1.3
4.3.1.3 If the
“Power On”
indication is not displayed and the cooling fan (Blower B1) is
not running, check to see that the unit is plugged into a live outlet and that the AC line
cord is plugged securely into the unit.
4.3.1.4
Check the output voltages from PS3; these voltages should be as follows:
PS3 J2, Pin 1
+12.0 ± 0.3VDC
PS3 J2, Pin 2
+ 5.0 ± 0.2VDC
PS3 J2, Pin 6
-12.0 ± 0.3VDC
If output voltages are not present on PS3, check the AC input to PS3.
4.3.1.5
Check the voltages to the A4 Operate/Control Board on connector A4 J3; the voltages
should be as follows:
A4 J3, Pin 16
-12.0
±
0.3VDC
A4 J3, Pin 29
+ 5.0
±
0.2VDC
A4 J3, Pin 31
+12.0
±
0.3VDC
4.3.1.6
Check the voltage on A4 J3, Pin 3; it should be
≥
4V when the
POWER
switch (S1) is in
the normal position and
<
0.1V when S1 is depressed. S1 is normally open; it is closed only
when it is depressed. The amplifier should change state every time the
POWER
switch is
depressed.
4.3.1.7
If all voltages are correct and the unit still does not operate, contact Amplifier Research to
arrange for repair or replacement of the A4 Operate/Control Board.
4.3.2 The Unit Cannot be Operated Remotely
4.3.2.1
Verify that the front panel
FUNCTION
switch is set to the
REMOTE
position.
4.3.2.2
Verify that the unit operates locally by resetting the
FUNCTION
switch to the
LOCAL
position; if the unit does not operate locally, see subsection
4.3.1
of this manual.
4.3.2.3
Check the position of the
“ADDRESS”
switch assembly (SW1) on the A4
Operate/Control Board; this assembly can be accessed through the unit’s rear panel. Check
to see that these switches are properly set for either RS-232 or IEEE-488 operation, as
desired.
(See subsection
2.2.2
of this manual for the proper
“ADDRESS”
switch settings.)
(
Note:
Address switches are only read at unit power-up; remove and re-apply AC power
(i.e., reset the circuit breaker) after changes are made.)