Model 150A400
Manual Text
4-4
REV B
4.3.3 Thermal Fault (Schematic Diagram No 1010877)
(continued)
4.3.3.1
Try to reset the unit; if the unit resets and operates normally, check to see that the cooling
fan (B1) is operating normally and that the air inlet on the bottom of the unit and the air
outlets on the rear of the unit are not blocked.
4.3.3.2
If the unit does not reset and the cooling fan is operating normally, check the voltage at
the A4 Operate/Control Board, J3, Pins 42 and 43; it should be
≤
0.1V.
4.3.3.3
If the voltage on A4 J3, Pin 21 is high, check the thermal daisy chain through A2AS1 and
A2BS1 to ground.
4.3.4 Interlock Fault (Schematic Diagram No. 1010877)
The Model
150A400
is equipped with an interlock connector, which is located on the rear panel. The
interlock circuit can be used to sense the openings of doors to screen rooms, test chambers, and so forth,
and to turn off RF energy when these doors are opened.
Note: The Model 150A400 is shipped with a mating connector, which has a jumper between
Pins 1 and 8, installed in the rear panel interlock connector. The unit will not operate unless
the interlock circuit is closed.
4.3.4.1
In the event of an Interlock Fault, the unit’s front panel VFD should indicate
“INTERLOCK
FAULT.”
4.3.4.2
Check to see if it is safe to power up the unit—are there personnel present in the screen
room, or are doors to the screen room open?
4.3.4.3
After checking for safety, try to clear the Interlock Fault from the front panel by using the
RESET
switch.
4.3.4.4
If the Interlock Fault will not clear, check for continuity in the External Interlock Circuit
(Pin 1 to Pin 8 in the connector, which mates with J2 in the rear panel).
4.3.4.5
Check the voltage on A4 J3, Pin 5; it should be
≤
0.1V.
4.3.4.6
If all of the above voltages are correct and the unit still will not reset, check for defective
wiring and/or PWB connections, then try the
RESET
switch again. If the unit still will
not reset, the A4 Operate/Control Board is defective. Contact Amplifier Research to
arrange for repair or replacement of the A4 Operate/Control Board.