21
Alarm Read Screen
The alarm read screen shows the configuration of the system
alarms. The second line of the display shows how often alarms will
be retransmitted. The third line displays current active alarm.
Note: where multiple alarms are active the screen will briefly rotate
each alarm into view.
Thus the following could appear on the third line display:
Temp:
Temp:
CAL ORIG
OUT OF CAL
Cond:
Cond:
CAL ORIG
OUT OF CAL
ORP:
ORP:
CAL ORIG
OUT OF CAL
pH:
pH:
CAL ORIG
OUT OF CAL
Conductivity
300s
ORP
300s
pH
300s
Sensor Alarms have a five minute delay. The 300s (Seconds) will
count down to zero before the alarm is activated.
Sensors Status Screen
This screen displays the status of sensors.
FLOW:
ON/OFF
CHEMS:
LO/OK
TEMP:
Current Temperature of the System (in oC)
DATALOG:
The Interval set for Logging
Conductivity Read Screen
This screen displays the conductivity status.
RL (relay):
OFF/ON
mS/cm:
millisiemens /centimetre
SET:
the current control set value for Conductivity
FLOW:
ON/OFF
ALM:
Will flash a value if alarmed
LOCK:
Not available on Conductivity
For descriptions and information on setting these items, refer to the
Set Conductivity Screen.