![Aquaread BlackBox Скачать руководство пользователя страница 60](http://html.mh-extra.com/html/aquaread/blackbox/blackbox_instruction-manual_2960649060.webp)
Aquaread BlackBox™ Instruction Manual
10401-00618 Rev P
8.4.22.
AP-7000
Sequence initiated with Start Measurement command:
Command
Response
<a>M!
or
<a>MC!
<a>0009<CR><LF>
<a>D0!
<a><baro><temp><ph><orp><cond><CR><LF>
or
<a><baro><temp><ph><orp><cond><crc><CR><LF>
<a>D1!
<a><cond20><cond25><res><sal><CR><LF>
or
<a><cond20><cond25><res><sal><crc><CR><LF>
Sequences initiated with Additional Measurement command:
Command
Response
<a>M1!
or
<a>MC1!
<a>0008<CR><LF>
<a>D0!
<a><tds><ssg><do><do-sat><aux1><CR><LF>
or
<a><tds><ssg><do><do-sat><aux1><crc><CR><LF>
<a>D1!
<a><aux2><aux3><aux4><CR><LF>
or
<a><aux2><aux3><aux4><crc><CR><LF>
Command
Response
<a>M2!
or
<a>MC2!
<a>0004<CR><LF>
<a>D0!
<a><aux5><aux6><nh3><depth><CR><LF>
or
<a><aux5><aux6><nh3><depth><crc><CR><LF>
Command
Response
<a>M3!
up to
<a>M9!
or
<a>MC3!
up to
<a>MC9!
<a>0000<CR><LF>
Sequence initiated with Start Concurrent Measurement command:
Command
Response
<a>C!
or
<a>CC!
<a>00020<CR><LF>
<a>D0!
<a><baro><temp><ph><orp><cond><cond20><cond25><res><sal><tds>
<ssg><CR><LF>
or
<a><baro><temp><ph><orp><cond><cond20><cond25><res><sal><tds>
<ssg><crc><CR><LF>
<a>D1!
<a><do><do-sat><aux1><aux2><aux3><aux4><aux5><aux6><nh3>
<CR><LF>
or
<a><do><do-sat><aux1><aux2><aux3><aux4><aux5><aux6><nh3><crc>
<CR><LF>
© 2019 Aquaread Ltd.
www.aquaread.com
Page 60 of 71