Aquaread BlackBox™ Instruction Manual
10401-00618 Rev P
8.4.4. AP-200 with Depth
Sequence initiated with Start Measurement command:
Command
Response
<a>M!
or
<a>MC!
<a>0004<CR><LF>
<a>D0!
<a><baro><temp><turb><depth><CR><LF>
or
<a><baro><temp><turb><depth><crc><CR><LF>
Sequences initiated with Additional Measurement command:
Command
Response
<a>M1!
up to
<a>M9!
or
<a>MC1!
up to
<a>MC9!
<a>0000<CR><LF>
Sequence initiated with Start Concurrent Measurement command:
Command
Response
<a>C!
or
<a>CC!
<a>00004<CR><LF>
<a>D0!
<a><baro><temp><turb><depth><CR><LF>
or
<a><baro><temp><turb><depth><crc><CR><LF>
Sequences initiated with Additional Concurrent Measurement command:
Command
Response
<a>C1!
up to
<a>C9!
or
<a>CC1!
up to
<a>CC9!
<a>00000<CR><LF>
Sequences initiated with Continuous Measurement command:
Command
Response
<a>R0!
or
<a>RC0!
<a><baro><temp><turb><depth><CR><LF>
or
<a><baro><temp><turb><depth><crc><CR><LF>
Command
Response
<a>R1!
to
<a>R9!
or
<a>RC1!
to
<a>RC9!
<a><CR><LF>
or
<a><crc><CR><LF>