8.4.15.
AP-800(software version 3.53 and below) and AP-900
Sequence initiated with Start Measurement command:
Command
Response
<a>M!
or
<a>MC!
<a>0009<CR><LF>
<a>D0!
<a><baro><temp><ph><orp><turb><CR><LF>
or
<a><baro><temp><ph><orp><turb><crc><CR><LF>
<a>D1!
<a><cond><cond20><cond25><res><CR><LF>
or
<a><cond><cond20><cond25><res><crc><CR><LF>
Sequences initiated with Additional Measurement command:
Command
Response
<a>M1!
or
<a>MC1!
<a>0005<CR><LF>
<a>D0!
<a><sal><tds><ssg><do><do-sat><CR><LF>
or
<a><sal><tds><ssg><do><do-sat><crc><CR><LF>
Command
Response
<a>M2!
up to
<a>M9!
or
<a>MC2!
up to
<a>MC9!
<a>0000<CR><LF>
Sequence initiated with Start Concurrent Measurement command:
Command
Response
<a>C!
or
<a>CC!
<a>00014<CR><LF>
<a>D0!
<a><baro><temp><ph><orp><turb><cond><cond20><cond25><res><sal>
<tds><CR><LF>
or
<a><baro><temp><ph><orp><turb><cond><cond20><cond25><res><sal>
<tds><crc><CR><LF>
<a>D1!
<a><ssg><do><do-sat><CR><LF>
or
<a><ssg><do><do-sat><crc><CR><LF>
Sequences initiated with Additional Concurrent Measurement command:
Command
Response
<a>C1!
up to
<a>C9!
or
<a>CC1!
up to
<a>CC9!
<a>00000<CR><LF>