Aquaread BlackBox™ Instruction Manual
10401-00618 Rev P
The BlackBox's sensor address is set to “0” during production. When a Change Address
command is received, the new address will replace the sensor address in the BlackBox’s
EEPROM and the BlackBox will only respond to subsequent commands (again with the
exception of the Address Query command) sent to that new address. The BlackBox will
always respond to the Address Query command, regardless of the stored sensor address.
8.3. Unit Identification
In its response to a Send Identification command, the BlackBox will report compatibility with
SDI-12 version 1.3, the vendor ID “AQUAREAD”, a sensor model number string dependent
upon the type of Aquaprobe connected to it (see Table 1), and the BlackBox's firmware
revision number and serial number.
Probe Type
Model Number String
AP-100
“AP100 ”
AP-100 with depth
“AP100D”
AP-200
“AP200 ”
AP-200 with depth
“AP200D”
AP-300
“AP300 ”
AP-300 with depth
“AP300D”
AP-400
“AP400 ”
AP-400 with depth
“AP400D”
AP-500
“AP500 ”
AP-500 with depth
“AP500D”
AP-600
“AP600 ”
AP-600 with depth
“AP600D”
AP-700
“AP700 ”
AP-700 with depth
“AP700D”
AP-800
“AP800 ”
AP-800 with depth
“AP800D”
AP-900
“AP900 ”
AP-900 with depth
“AP900D”
AP-1000
“AP1000”
AquaPlus
“AQ+ ”
AP-2000
“AP2000”
AP-2000 with depth
“AP2KD”
AP-Lite
“APLITE”
AP-5000
“AP5000”
AP-7000
“AP7000”
Table 1: Probe Type Model Number Strings
8.4. Measurement Commands
On reception of any Start Measurement, Additional Measurement, Start Concurrent
Measurement or Additional Concurrent Measurement command (with or without CRC
request), the BlackBox will respond indicating a measurement time.
If the BlackBox is configured in ‘always on’ mode, the measurement time will be “000”. If
the BlackBox is configured in ‘auto sleep’ mode, the measurement time can be up to “040”.
Following the measurement time, the BlackBox will send the number of measurements it
will be sending dependent upon the type of Aquaprobe connected to it.
If the measurement time sent by the BlackBox is “000”, the data recorder should
immediately respond with a Send Data command, to which the BlackBox will respond with
a packet containing appropriate measurements dependent upon the type of Aquaprobe
connected to it and upon the measurement set requested. The response will also contain a
CRC, if requested.