Aquaprobe
®
AP-5000 Instruction Manual
10401-00906 Rev M
Now that this GS Factor has been applied to the electrode, all future measurements will be
multiplied by 1.25 prior to being displayed.
In this way, the electrode has been corrected for the local conditions and species of organic
matter.
15.11.8.
Errors During Calibration
If a problem occurs during calibration, an error message will be displayed. Refer to
in section 10 for error message handling.
15.11.9.
Lens and Sleeve Maintenance
On a daily basis, the lenses on the electrode should be wiped over with a soft damp cloth.
Similarly, the inside of the Probe Sleeve and Sleeve Cap should be kept clean and free
from any deposits that may cause stray fluorescence.
Never use an abrasive cleaner on the inside of the Probe sleeve or cap as they have
been treated with a non-reflective coating which can be easily damaged.
The inside of
the sleeve should be wiped over with a soft damp cloth and
non-abrasive
detergent.
Always re-calibrate the zero point after cleaning the sleeve or lenses.
15.11.10.
Special Note Concerning CDOM Calibration Solutions
All types of CDOM calibration solution are acidic and will therefore, given time, attack the
anodised finish on the Aquaprobe.
In order to avoid this, do not leave the Aquaprobe sitting in CDOM solution for more time
than it takes to do the actual calibration.
As soon as the Point 2 calibration is complete, remove the Aquaprobe from the CDOM
calibration solution and rinse thoroughly.