Aquaprobe
®
AP-5000 Instruction Manual
10401-00906 Rev M
During long term deployment of ion selective electrodes, the user should obtain grab
samples during the course of the deployment for analysis in the laboratory by
chemical means and use the results to apply post calibration to the recorded results.
5.
Accuracy in the field is dependent upon full three-point calibration and minimal time
between calibration and use.
6.
In order to achieve accurate readings with ISE electrodes, the Probe needs to be
either placed in flowing water, or needs to be stirred or raised and lowered
continuously to ensure a minimum flow rate of 0.3m/s over the electrode.
If there is
no water flow across the ISE electrode, the ions in the immediate area of the
electrode will be depleted and the reading will start to fall. This also applies to
calibration, where the probe should be stirred at all times.
25.1. Special Notes Concerning ISE Electrodes and pH Buffers
The high ionic concentration of pH calibration
solutions (buffers), including RapidCal, can
cause significant offsets in ISE electrodes.
These offsets are temporary, but best avoided
because they can cause significant errors
during both calibration and normal operation.
For this reason all ISE electrodes are supplied
with a red rubber sealing cap.
The caps should be fitted to all ISE
Electrodes during pH calibration or when
using RapidCal
in order to protect the ISE
electrodes from the effects of the buffer
solution.
At all other times, the ISE electrodes should be left uncovered.
© 2019 Aquaread
®
Ltd.
www.aquaread.com
Page 131 of 138