22.2. Socket Assignment and Calibration
After installation, it is essential to connect the AP-5000 to an Aquameter
®
and assign the
new electrode types to the relevant AUX Sockets. On the Aquameter
®
, press the MENU
key, then select Setup & Install followed by Socket Assignment. When the Socket
Assignment option has been selected, the following screen will be displayed.
SOCKET ASSIGNMENTS
1:EMPTY
l
4:EMPTY
2:EMPTY
l
5:N/A
3:EMPTY
l
6:N/A
Using the up and down arrow keys, select the AUX socket you wish to assign then move
the cursor to the right by pressing the right arrow key. When the cursor has moved to the
right of the AUX socket number, use the up and down arrow keys to select the appropriate
electrode type. The tables below show the available electrode options and the selection
that should be made on this screen:
AP-2000/5000 Optical Electrodes
Electrode Part No.
Function
Aquameter
®
Selection
2000/5000-TURB
Turbidity
TURB
2000/5000-CPHYLL
Chlorophyll
Cphl
2000/5000-BGA-PC
Phycocyanin (Blue-Green Algae PC)
BGA-PC
2000/5000-BGA-PE
Phycoerythrin (Blue-Green Algae PE)
BGA-PE
2000/5000-RHOD
Rhodamine WT Dye
Rhod
2000/5000-FSCEIN
Fluorescein Dye
Fcein
2000/5000-REFOIL
Refined Oil
R-OIL
2000/5000-CDOM
CDOM/FDOM
CDOM
AP-5000/7000 ISE Electrodes
Electrode Part No.
Function
Aquameter
®
Selection
5000/7000-AMM
Ammonium/Ammonia
NH4
5000/7000-CHL
Chloride
Cl
5000/7000-FLU
Fluoride
F
5000/7000-NIT
Nitrate
NO3
5000/7000-CAL
Calcium
Ca2
When the desired electrode type is showing, move the cursor back to the left of the socket
number then press OK to send the selection to the AP-5000. The socket assignments are
stored in the AP-5000. If you press the ESC key whilst in this screen, any changes you
have made will not be transferred to the AP-5000.
Finally, refer to the relevant section of this manual and carry out a full two-point (optical) or
three-point (ISE) calibration of the new electrode.
YOUR NEW ELECTRODE WILL NOT GIVE SENSIBLE READINGS UNTIL IT HAS BEEN
FULLY CALIBRATED.
Please note: changing an AUX Socket assignment will clear all the calibration data
for that socket.
Keep the blanking plug in a safe place. If you subsequently remove an electrode, be sure to
replace the blanking plug and set the socket assignment back to EMPTY.