Aquaprobe
®
AP-5000 Instruction Manual
10401-00906 Rev M
6. Select
Full Cal.
The screen will change to:
Calibration
pH/REDOX(ORP)
DO/EC
Aux Electrodes
7. Select
pH/ORP
. The screen will change to:
pH 7.00?[01/Jan/17]
PH 4.01?[01/Jan/17]
PH 10.0?[01/Jan/17]
ORP+?250[01/Jan/17]
The dates shown to the right of the screen are the dates of the last successful calibration.
8. Select pH7.00. The screen will change to:
PLEASE WAIT
Stabilising
000%
The Meter will wait until the readings are stable, then it will send the calibration command
to the Probe, where the calibration takes place. During calibration, the Calibrating screen is
displayed and the progress counter counts up. If the calibration is successful, the counter
will reach 100% and the following screen will be displayed.
Offset: -1.2mV
Calibrating
100%
Press [OK]
The top line displays the voltage offset from zero for the pH electrode in +/-millivolts (mV). If
this offset goes /-25mV at 25
°
C, the pH electrode should be serviced.
This value is stored in the Probe's memory and can be recalled at any time. See section
10.7. Calibration Data Storage and Retrieval
Press OK then ESC repeatedly to return to
normal reading mode.
Remove the Probe from the calibration cup, rinse thoroughly in de-ionised water, shake off
any excess and dry the outer sleeve with a soft cloth.