![Aquaread Aquaprobe AP-5000 Скачать руководство пользователя страница 68](http://html.mh-extra.com/html/aquaread/aquaprobe-ap-5000/aquaprobe-ap-5000_instruction-manual_2960643068.webp)
Aquaprobe
®
AP-5000 Instruction Manual
10401-00906 Rev M
15.7.6.
Zero Point Calibration
To calibrate the zero point, follow these steps:
1. Pour 300mL of clean water (bottled still mineral water is recommended) into a clean
calibration cup, remove the storage cap from the pH electrode if fitted, wash the
Probe in clean water, then gently lower the Probe in all the way.
The Sleeve End
Cap must be fitted
.Bang the probe against the bottom of the cup several times to
dislodge any air bubbles.
2. Wait until the temperature and Cphl readings are stable. If the Cphl reading is very
high, there are probably air bubbles adhering to the lenses.
3. Ensure the temperature of the solution is between 5ºC and 40ºC (41ºF - 104ºF).
4. Press the
MENU
key then select
Calibration.
The following screen will be displayed.
Calibration
RapidCal
DO 100%
Full Cal
5.
Select
Full Cal.
The screen will change to:
Calibration
pH/REDOX(ORP)
DO/EC
Aux Electrodes
6.
Select
Aux Electrodes
. The screen will change to:
SELECT ELECTRODE
1:Cphl
l
4:EMPTY
2:EMPTY
l
5:N/A
3:EMPTY
l
6:N/A
The Cphl electrode should have been assigned to an AUX socket when it was fitted.
Choose that socket. Press the OK or right arrow key to select Cphl. The screen will change
to:
CALIBRATE Cphl
ZERO? [01/Jan/17]
Pt-2? [01/Jan/17]
GS Factor:01.00
Calibration point 2 (Pt-2) is the upper calibration point.
The dates shown to the right of each point are the dates of the last successful calibration.
© 2019 Aquaread
®
Ltd.
www.aquaread.com
Page 68 of 138